MULTICAL® 803
Kamstrup A/S · Technical description · 5512-2360_A1_GB_02.2019
57
3.5
Data
The country code is selected as the last two characters of the calculator’s static part of the type number.
Static data
803-xxxx
-
Dynamic data
xxxx
Written on the
calculator’s front
Can be
displayed
Type 803 -
□
-
□
- XX -
□
-
□□
-
□
In addition to being used to define the language on the meter label and the approval and verification labels, the
country code is used to control the configuration of certain meter data.
During production, MULTICAL® 803 is programmed with a number of measuring values in the fields listed overleaf.
Unless the order includes specific requirements as to configuration, MULTICAL® 803 is supplied with the default
values from the table.
Automatic
To be stated in order
Default
Serial no. (S/N)
1
79,800,000
-
-
Customer no.
Display No. 1 = 8 digits MSD
Display No. 2 = 8 digits LSD
-
Up to 16 digits
Customer no. equal to
S/N
Yearly target date 1 (MM.DD)
-
MM=1-12 and DD=1-31
Depends on country
code
Monthly target date 1 (DD)
-
DD = 1-31
Yearly target date 2 (MM.DD)
-
MM=1-12 and DD=1-31 +
00.00 (disabled)
2
Monthly target date 2 (DD)
-
DD = 1-31 +
00 (disabled)
2
Averaging period 1 of min./max of
power (P) and flow (Q) for yearly
and monthly loggers
(See paragraph 7.5)
-
1…1440 minutes
60 minutes
Averaging period of min./max
temperature (t) for yearly and
monthly loggers
(See paragraph 7.5)
5 min.
-
Depends on country
code
Averaging period of CP
(See paragraph 7.2)
-
5…30 days
7 days
θ
Hc
Heat/cooling shift
Only active if meter type 6 is
selected
(See paragraph 7.4)
-
2…180.00
°
C + 250.00
°
C
3
Heat/cooling, meter type
6:
25.00
°
C
Date/time
20YY.MM.DD/
hh.mm.ss
GMT
±
12.0 hours
(Can be defined at half hour
intervals)
-
GMT offset
-
-
Depends on country
code