Full Digital Color Doppler Ultrasonic Diagnostic Instruments (Vet) User’s Manual V1.01
- 76 -
14.1.2 Clean the probe cable and its connector
1. Clean the probe cable and its connector with soft, dry cloth.
2. In case of die-hard blots, clean with soft cloth dipped in moderate detergent and then air-dry it.
14.1.3 Clean the LED screen
Use a soft cloth dipped in glass cleaner to clean the LED screen, and then air-dried.
14.1.4 Clean the control panel, shell and probe bracket
Clean the instrument surface with soft, dry cloth or with soft cloth dipped in moderate water cleaning
media to remove the blots, and then dry the instrument with soft, dry cloth or with air.
14.2 Replace the fuse
The type of the fuse used in this system is 5HTP3.15-R (250V~, T3.15AH). The fuse should be
replaced by the service personnel or the designated technical personnel.
14.3 Replacement of power supply cord
Before replacing the power supply cord, please contact us; replace the power supply cord under the
guidance of our company. Please use the power supply cord provided by us.
14.4 Using the Battery
14.4.1 Precautions
1. Battery is consumable; the battery cycle-life is based on the times of charge and discharge as unit.
When the use time reduced significantly compared with normal conditions, the battery should be
promptly replaced.
2. The excess high or low temperature will affect the charging and discharging performance, and short
the battery life and capacity.
Attention:
Do not clean the screen with hydrocarbon detergent for example alcohol etc
or OA equipment cleaning media.
Attention:
Prohibit using sharp objects to touch the LED screen, and prohibit pressing
or squeezing against the LED screen.
Danger:
1.
Do not assemble or disassemble the battery, because there is danger of explosion.
2.
Charge the battery through the diagnostic ultrasound system only. Otherwise, there is
danger of explosion.
3.
Replace the battery provided by the manufacturer only, and can only be replaced by
technical personnel.