4.
Move the marker to the starting point of peak waveform of the end systole by using the trackball, press
Set
to fix the marker and the second marker displays.
5.
Move the second marker to the end diastole which is one cardiac cycle adjacent to the first marker by
using the trackball and press
Set
. The system automatically traces the waveform and the measurement
result displays in the result box.
10.6 Specific Measurements and Calculations
Specific measurement and calculations corresponds to the measurement and calculation of each exam type.
The measurement results are recorded in the report.
You should confirm the following items before measurement:
Confirm that the current probe is applicable for the current application part.
Confirm that the current date is the correct.
Confirm that the patient information is input.
Confirm that the exam type is selected.
10.6.1 Abdomen Measurements
Abdominal measurements can be performed in B/Color/PDI mode.
Steps:
1.
Press the
Measure
key in the real time or frozen B/Color/PDI mode, and the measurement menu
displays on the top left of the screen.
2.
Select a measurement item on the menu by moving the trackball, such as [Liver], and press
Set
to
confirm.
3.
Select the measurement sub-item on the pop-up sub-menu, such as [R Liver Max], and press
Set
, a
marker displays on the image.
4.
Follow the methods in the following table to perform the measurement, and the measurement result
displays in the result box.
Measurement Item
Measurement Sub-item
Measurement Method
Liver
R Liver Max
Refer to Section 10.5.1.1 Distance.
R Liver W
R Liver D
L Liver L
L Liver D
Gallbladder
(GB)
CBD
Refer to Section 10.5.1.1 Distance.
CHD
GB L
GB W
GB Wall D
Pancreas
Pancr. Head
Refer to Section 10.5.1.1 Distance.
Pancr. Corus
Pancr. Tail
Pancr. Duct
Spleen
SP L
Refer to Section 10.5.1.1 Distance.
SP D
SP W
Right kidney
RK L
Refer to Section 10.5.1.1 Distance.
RK D