Chapter 9 Processing the Images
You can work with the acquired images by using the features provided by the ultrasound system, such as
reviewing, magnifying and freezing the image, cropping the cine, and making annotations on the image.
9.1 Displaying the image
9.1.1 Dual-splitting Display
You can position two images side by side on the screen to compare the images acquired at different times.
Steps are as follows:
1.
Press B, Color or PDI on the control panel to enter the real time scan.
2.
Press
to enter the dual-splitting display.
The image is spitted by two parts with a real time scan on
the left and no image on the right.
3.
Press
again to freeze the real time scan and unfreeze the frozen image.
The probe direction mark is highlighted in the a real time scan and becomes gray in the frozen mode.
4.
Press
in turn will freeze the current real-time image, the other image gets into real-time status.
Press other imaging mode keys to exit the dual-splitting display.
9.1.2 Quad-splitting Display
You can position four images side by side on the screen to compare the images acquired at different times.
Steps are as follows:
1.
Press B, Color or PDI on the control panel to enter the real time scan.
2.
Press
to enter the quad-splitting display. The image is spitted by four parts with a real time scan
on the top-left and no image on the rest parts.