background image

JPD-100S9

Fetal Doppler

USER MANUAL

Copyright © Jumper Medical. All rights reserved.
Manual Version:Version 1.1
Issuing Date: 2020.11

Содержание AngelSounds JPD-100S9

Страница 1: ...JPD 100S9 Fetal Doppler USER MANUAL Copyright Jumper Medical All rights reserved Manual Version Version 1 1 Issuing Date 2020 11...

Страница 2: ......

Страница 3: ...s Email info jumper medical com Website http www jumper medical com MedPath GmbH Mies van der Rohe Strasse 8 80807 Munich Germany Please read the User Manual carefully to make sure safe and proper use...

Страница 4: ...arance 3 Controls and Indicators 4 Location and Detection 8 Recording your Baby s Heartbeat 10 Share your Recording on Social Media 11 Principles of how the Fetal Doppler works 11 Cleaning Disinfectin...

Страница 5: ...then troubleshoot before trying to find the fetal heartbeat 4 When the Fetal Doppler is working correctly you can now lay back and apply coupling agent to the surface of the ultrasonic probe 5 Now pu...

Страница 6: ...y s heartbeat The normal range of fetal heart rate 110bpm 160bpm CAUTION It cannot replace the professional fetal monitor when the fetal heart rate is abnormal or can not find the fetal heart can not...

Страница 7: ...ce used for detecting fetal heartbeats Note for home use This device cannot replace a professional fetal monitor If the fetal heart rate is abnormal or cannot be located using this monitor immediately...

Страница 8: ...d to change the volume Press the Volume up button the sound volume will increase Contrary press the Volume down button the sound volume will decrease Recording jack Used to connect a smartphone via a...

Страница 9: ...and matched 3 Close the battery cover as the bellow picture shows Additional Battery Notes Be sure to remove the battery s protective clear wrapper before installing it in the Fetal Doppler If the Fe...

Страница 10: ...tore or Google Market by entering the keyword JUMPER Health Notes This APP supports iOS 7 0 and later versions and Android 4 3 and later versions In addition hardware of the smartphone needs to suppor...

Страница 11: ...on the belly Baby Oil Coconut Oil Virgin Olive Oil or even Vaseline work well which is also readily available in most homes Actually anything that lubricates the movement across your belly will work...

Страница 12: ...omen to listen to the baby s heartbeat which should be much faster than mommy s heartbeat 6 Move the Fetal Doppler very very slowly to find the optimum placement based on the baby s current position 7...

Страница 13: ...a break and trying again a few days later However if you are concerned about any issues relating to your pregnancy you should contact your healthcare provider for assistance As the baby grows the hea...

Страница 14: ...Fetal Doppler is a listening system to hear the sounds your baby makes and is NOT a medical device and can not be used for diagnostic or medical purposes and is not to be used as a substitute for regu...

Страница 15: ...ces off of the baby s heart The difference between these two signals sound waves are measured to determine the baby s heartbeat Cleaning Disinfecting Care and Disposal We recommend cleaning the gel fr...

Страница 16: ...eak battery can cause noise Interference can be caused by nearby electronic devices e g high frequency electronics cell phones cordless phones etc Make sure there is plenty of coupling gel between the...

Страница 17: ...sier to hear the heartbeat when you find the right location The baby can sometimes be located behind the placenta making it difficult to find try again later The battery cover doesn t want to stay on...

Страница 18: ...equipment P_ 1MPa Iob 20mW cm2 Ispta 100mW cm2 Audio output power 0 2 W Audio out jack 3 5mm Recommended battery type 9V DC alkaline battery Stand by Time 4hours Suitable gestation 16 gestational wee...

Страница 19: ...tion against ingress of water and particulate matter It indicates that the equipment should be sent to the special agencies according to local regulation for separate collection after its useful life...

Страница 20: ...suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domesticpurposes A...

Страница 21: ...in the electromagnetic environment specified below The customer or the user of the Fetal Doppler should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance leve...

Страница 22: ...rption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM...

Страница 23: ...er of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7...

Страница 24: ...Shenzhen Jumper Medical Equipment Co Ltd Address D Building No 71 Xintian Road Fuyong Street Baoan Shenzhen Guangdong China Tel 86 755 26696279 Fax 86 755 26852025 Website http www jumper medical com...