8 Programming
96
8.6.3 Interface
Sets the parameters of the RS 422/RS 485
interface:
- protocol
- baud rate
- data format
- instrument address
- minimum response time
The minumum response time is the mini-
mum time span between the query of an in-
strument in a data network and the answer
of the pen recorder.
Extra Code 52 or 53 is required.
B 70.6001.2
Display
Edit
Selection/input
with keys
Enter
Continue with
INTERFACE
E
PROTOCOL: J-BUS
Set protocol:
J-BUS
MOD-BUS
h
,
R
E
2
2
BAUD: 9.6kBAUD
Select baud rate
(values in kbaud):
0.15, 0.3, 1.2, 2.4, 4.8,
9.6, 19.2, 38.4, 125.0,
187.5kbaud
h
,
R
E
3
3
DATA FORMAT:
8/1/NO
Select data format
(data bits/stop bits/
parity):
8/1/NO,
8/1/ODD,
8/1/EVEN,
8/2 NO,
8/1/ZERO
h
,
R
E
4