background image

7 Text printing

36

7.4   Scaling

The scaling can be printed in two ways: 

cyclic:

for each channel at a configured
spacing

triggered: for all channels on pressing the

(

F

 key (hold down for at least 4

sec) or by closing one of the log-
ic inputs.

Cyclic scaling print-out

Spacing which can be selected:

- 30 cm approx.

- 60 cm approx.

- 90 cm approx.

- off

The lowest line shows the channel number,
the corresponding pen colour, the pro-
grammed channel designation and the unit
of the measured value.

If the range 0 —100mm has not been al-
tered by offset, the next line shows the scale
values for 0%, 50% und 100%. If the range
has been restricted, this line contains marks
for the start and the end of the range. The
corresponding numbers are printed in the
next line.

The limit marks for the current channel are
printed in the top line.

Содержание LOGOLINE 500d

Страница 1: ...LOGOLINE 500d Pen recorder with text printing and LED dot matrix display B 70 6021 0 Operating Manual 2011 01 19 00331468...

Страница 2: ...for special function hold down at least 4sec factory setting scale printout l r Cursor left right select digit l r Deactivate activate display of status error messages h R Select display date and time...

Страница 3: ...4 1 Notes on installation 15 4 2 Connection diagram 16 5 Starting up 18 5 1 Display and controls 18 5 2 Opening and closing the door 18 5 3 Marking the channel label 18 5 4 Fitting the fibre pens 19 6...

Страница 4: ...4 6 Plotarea zoom 76 8 4 7 Presentation range offset 77 8 5 Configuration level 2 78 8 5 1 Instrument designation 80 8 5 2 Speed programming mode 81 8 5 3 Speed limit operation 82 8 5 4 Timed operatio...

Страница 5: ...10 1 Converting the chart cassette 110 10 2 Extra Code 259 111 10 3 External relay module ER8 112 10 4 Setup program 115 10 5 Maths and logics module 117 10 6 Interface RS422 RS485 118 11 Fault findi...

Страница 6: ...tions inside the instrument are described in this Operating Manual However if any difficulties should arise during start up please do not carry out any manipulations You could endanger your rights und...

Страница 7: ...ition to installation and electrical con nection it contains all the information on in strument start up operation and parameter setting Interface Description B 70 6001 2 This Operating Manual is supp...

Страница 8: ...s disregarded or not followed accurately E Warning This symbol is used where precautions have to be observed during the handling of components which may be damaged by electrostatic discharges 1 3 2 No...

Страница 9: ...mbols and text are possible Where a key has mul tiple functions the text shown is the one corresponding to the actual fuctions dis cussed Program manager Screen texts Texts displayed in the setup prog...

Страница 10: ...data are stored permanently in EEPROM Input signals include resistance thermome ters resistance transmitters thermocouples standard current and voltage signals The necessary linearisation is performe...

Страница 11: ...GOLINE 500d with 1 universal input and text output 706022 LOGOLINE 500d with 2 universal inputs and text output via channel 1 706023 LOGOLINE 500d with 3 universal inputs and text output via channel 1...

Страница 12: ...or tension relief of the sensor leads connected 1 Fibre pen disposable for each channel 2 Chart rolls 16m long or 1 Chart roll 32m long with Code r32 or 1 Fanfold chart pack 16m long with Code fp 2 3...

Страница 13: ...ion Stray electromagnetic fields e g from mo tors transformers etc should be avoided where possible The ambient temperature at the location may be between 10 and 50 C at a relative humidity not exceed...

Страница 14: ...back of the panel hook the mounting brackets 1 into the cut outs in the sides of the housing The flat bracket faces must lie against the housing Place the brackets against the rear of the panel and ti...

Страница 15: ...tional loads to the supply terminals of the instrument If contact with live parts is possible when working on the instrument it has to be isolated on both poles from the supply The instrument is not s...

Страница 16: ...erly qualified personnel Terminals Supply as on label N neutral L1 line PE potential earth N L L1 L PE Analogue inputs Input 1 Input 2 Input 3 Field Field Field Voltage input up to 170 mV 1 2 3 Voltag...

Страница 17: ...t S slider E end Field External relay module ER8 Communication with ex ternal relay module ER8 8 Supply for external 2 wire transmitter 24V 5 45 mA Logic operating inputs Contact operation LOW ROFF 10...

Страница 18: ...ay 7 keys for operation and programming 8 door 9 housing to DIN 43700 for flush panel mounting galvanised steel 10 channel marker label 5 2 Opening and closing the door Turn the knob to open or close...

Страница 19: ...ng up 19 5 4 Fitting the fibre pens Open the recorder door Stop recording e g press key S Swing display upwards Slide the fibre pen into the holder up to the stop Swing display down again until it cli...

Страница 20: ...ding is stopped and continued using key S In the stop status the display flashes STOP Stop via external stop External stop can be operated through a signal at a logic input on the back of the recorder...

Страница 21: ...by closing a logic input Section 7 4 Section 8 4 3 Plotarea During plotarea operation a portion of the total measur ing range of a trace is recorded to a larger scale Section 8 4 6 Section 10 4 Offset...

Страница 22: ...he para meter parameter level display time one channel and its measurement as a number or bargraph or system error messages The display can be altered during recording using the keys h and R If there...

Страница 23: ...odenumber the recorder interrupts signal acquisition and recording at this level Statistical reports are re started on leaving the parameter level The following parameters are altered at the parameter...

Страница 24: ...ters writing status signal input scaling channel designation limit operation plotarea zoom offset presentation range H If a correct codenumber has been input when calling up the parame ter level the r...

Страница 25: ...ing mode chart speed limit operation timed operation scale printing time printing pen offset compensation report start text end text pre setting codenumber H If a correct codenumber has been input whe...

Страница 26: ...al text binary linked external text external stop external speed event counter external scaling external report For a summary of all the parameters of the recorder see the Appendix H If a correct code...

Страница 27: ...us the program branches to the individual levels and from there into the corresponding parameters and also into any sub parameters To alter a certain parameter run through the relevant levels up to th...

Страница 28: ...ey functions e Chart fast forward when recording is stopped in basic status Aborting parameter input Exit Level change backwards S Start Stop recording Decimal place selection during value input For t...

Страница 29: ...ment and decrement the selected diget with the h and R keys shift the decimal point with the S key enter value input with E Five keys are used to input text select the character to be altered with the...

Страница 30: ...ich are as signed priorities in the setup program These priorities determine abort criteria where printing of several texts is required simultaneously The following applies in the explanation below te...

Страница 31: ...inted for this overlimit low priority as follows 12 01 relay 1 ON 12 02 relay 1 OFF and 12 03 relay 1 ON In this example the printing of the report would first be completed This is followed by the del...

Страница 32: ...est is started service print is started recorder is switched off An exception is the statistical report The following fixed priorities are allocated to stop service print and print test Section 7 9 Pr...

Страница 33: ...xt is typically 10 40sec Advantages quick text printing minimum blockage of other messages text printing also at chart speeds above 360 mm h text printing also when zero chart speed has been programme...

Страница 34: ...ed chart speed Advantages the time reference is retained during text printing recording of signals on channel 1 is only interrupted for a very brief time if pen offset compensation has been switched o...

Страница 35: ...d the spacings given above are only approximate Every fourth clock time print is followed alternately by the current chart speed the programmed recorder designation and the date The date is always pri...

Страница 36: ...ox 60 cm approx 90 cm approx off The lowest line shows the channel number the corresponding pen colour the pro grammed channel designation and the unit of the measured value If the range 0 100mm has n...

Страница 37: ...down for at least 4 sec Alternatively the triggered scaling print out can be started using a logic input The print out covers the scaling of all chan nels whose writing status is switched on Print ou...

Страница 38: ...eed timed operation The indivual events have different priority The current chart speed depends on the event with the highest priority Event Priority limit operation external speed timed operation nor...

Страница 39: ...corresponding text can not be printed This is the case when the setup pro gram has been configured under Edit Print out texts in Print mode Process value overwrite for Text on Change of Chart Speed an...

Страница 40: ...being printed in the stop status in which no other text can be printed there is therefore no priority and no trace interrupting overwriting for the end text The start text should always have the high...

Страница 41: ...s Following each change of status of these logic inputs the text corresponding to the binary number is printed Text printing can be switched off separately for each of the 16 texts Example If the stat...

Страница 42: ...n to the programmed text the time is printed at which the text printing has been requested The request for printing a relay text is ignored if the text consists only of blank characters There is also...

Страница 43: ...set using the setup program Each incrementing is reported by printing the clock time and the programmed text fol lowed by the current counter reading Section 8 6 8 Section 10 4 The time reference mark...

Страница 44: ...recorder is re configured from the keys or if the setup connector is plugged in If printing of the statistical report is inter rupted by stop re configuration setup or a print test or service print p...

Страница 45: ...keys It is used to check the function of writing system and fibre pens Started text printing is aborted and not continued after the print test All text printing requests arriving during the print test...

Страница 46: ...Started text printing is aborted and not continued after the service print All text printing requests arriving during the service print are rejected The only exception is the statistical report If the...

Страница 47: ...h or status error messages Display can be cyclic or static cyclic display After approx 3 sec the pen recorder switch es automatically to the subsequent display static display In the basic status the s...

Страница 48: ...8 Pressing the e key in the stop status ac tivates the chart fast forward Pressing the F key hold down at least 4 sec starts the scaling print out Pressing the h R keys simultaneously changes to the o...

Страница 49: ...essing the e key The operating level includes the following parameters chart speed print test service print A summary of all parameters of the recorder is given in the Appendix Section 12 6 While the...

Страница 50: ...et to standard chart speeds In the setup program the program ming mode can be changed to con tinuous under Edit Chart Speed For alteration on the instrument Section 8 5 2 Please note Section 7 2 Displ...

Страница 51: ...is terminated by PRINTTEST OFF Started text printing is aborted and is not continued after the print test All text printing requests arriving during the print test are ignored and rejected The only ex...

Страница 52: ...vice print All text printing requests arriving during the service print are ignored and rejected The only exception is the statistical report If the maths and logics module is activated an M is printe...

Страница 53: ...r Edit Instrument operation Parameters which are not required for the applications or which are only programmed once can be marked there accordingly Af ter the new set up has been transferred to the p...

Страница 54: ...g and recording remain activated The parameters can be viewed but not programmed The codenumbers can be altered in the setup program under Edit Codenumber or on the instrument Section 8 5 12 Display E...

Страница 55: ...me If an incorrect codenumber is en tered the recorder remains in nor mal operation Signal acquisition processing and recording remain activated The parameters can be viewed but can not be programmed...

Страница 56: ...nglish French German The language setting affects all non pro grammable texts in the recorder which are printed or indicated Display Edit Selection input with keys Edit Continue with LANGUAGE E Langua...

Страница 57: ...d summer time Date and time are checked for plau sibility after input An error message is output where appropriate The error message is acknowl edged with E and the data can then be re entered Display...

Страница 58: ...ne hour to 2 00 hrs The time interval for summer time always applies only to the period in dicated It has to be re programmed afresh for each year Display Edit selection input with keys Enter Continue...

Страница 59: ...8 3 4 Display brightness The brightness of the display can be pro grammmed in four different steps Display Edit Selection input with keys Enter Continue with DISPLAY E BRIGHTNESS STEP 1 4 h R E h forw...

Страница 60: ...selected also the limit comparator function switch on or off at overlimit or underlimit lk7 lk8 concerning lk7 und lk8 This parameter is only visible when the recorder is fitted with the inter face fo...

Страница 61: ...isplay when the recorder is in the basic status The time display can be suppressed status OFF by using this parameter The instrument version can be determined by the parameter Operating level Service...

Страница 62: ...nation limit operation plotarea zoom offset presentation range Following the input of a correct codenumber when calling up the pa rameter level the recorder interrupts signal acquisition and recording...

Страница 63: ...e is printed on the chart and appears in the statistical report or not Display Edit Selection input with keys Enter Continue with PLOTSTATE E 1 PLOTSTATE Select number of the recording channel h R E 2...

Страница 64: ...pied to the scaling after editing This does not apply if linear cus tomized table 1 or customized table 2 has been configured as line arisation Display Edit Selection input with keys Enter Continue wi...

Страница 65: ...3 wire 4 wire Define type of connection of resistance thermometer 2 3 wire or 4 wire h R E 32 32 1 UNIT C C F h R E 33 33 1 RANGE BEG 200 0 C Input start of measurement range h R l r S E Check range s...

Страница 66: ...Input end of measurement range h R l r S E Check range end according to characteristic end start 100 C S R B 500 C 44 44 1 TEMP COMP INTERN Pt 100 INTERN Pt 100 EXTERN Pt 100 EXT CONST h R E INTERN Pt...

Страница 67: ...FILTER otherwise 51 51 1 CONNECTION 2 3 wire 2 3 wire 4 wire Define type of potentio meter connection 2 3 wire circuit or 4 wire circuit h R E 52 52 1 Ro 300 0 Ohm Input start resistance h R l r S E C...

Страница 68: ...R E linear characteristic 61 customized X1 oder X2 10 FILTER 61 1 Ra 300 0 Ohm Input start resistance h R l r S E Check 0 start res 3900 Ohm 62 62 1 Rs 300 0 Ohm Input measurement range h R l r S E Ch...

Страница 69: ...20 C X1 customized lin 1 X2 customized lin 2 h R E customized X11 or X21 10 FILTER otherwise 71 71 1 UNIT mV mV V h R E 72 72 1 RANGE BEG 0 000 mV Input start of measurement range h R l r S E Check ra...

Страница 70: ...1 TEMP C C F h R E 75 75 1 TEMP BEG 200 0 C Input start of temperature range h R l r S E Check temp start according to characteristic 76 76 1 TEMP END 900 0 C Input end of temperature range h R l r S...

Страница 71: ...400 C K 270 to 1372 C E 270 to 1000 C N 270 to 1300 C S 50 to 1768 C R 50 to 1768 C B 0 1820 C X1 customized lin 1 X2 customized lin 2 h R E customized X11 or X21 10 FILTER otherwise 81 81 1 RANGE BEG...

Страница 72: ...se 83 83 only with ther mocouples and resistance thermometers 1 TEMP C C F h R E 84 84 1 TEMP BEG 200 0 C Input start of temperature range h R l r S E Check temp start according to characteristic 85 8...

Страница 73: ...nge cor responds to a temperature of 200 600 C Programming Unit C Scaling start 200 Scaling end 600 Number format XXX X Section 7 4 Display Edit Selection input with keys Enter Continue with SCALING E...

Страница 74: ...shown on the display together with the measurement value The complete character set is availa ble v Section 12 5 Display Edit Selection input with keys Enter Continue with WORD AND NUMBER E 2 WORD AND...

Страница 75: ...o frequently The switching differential is referred to scaling and plotarea zoom Display Edit Selection input with keys Enter Continue with LIMIT FEED E 2 LIMIT FEED Select recording channel h R E 2 2...

Страница 76: ...ed as follows The difference between low and high value must be at least 10 Display Edit Selection input with keys Enter Continue with PLOTAREA E 2 PLOTAREA Select recording channel h R E 2 2 2 LOW 01...

Страница 77: ...lected for the display Sets the start and end of the trace on the chart The input is made in mm steps Display Edit Selection input with keys Enter Continue with OFFSET E 2 OFFSET Select recording chan...

Страница 78: ...key The configuration level 2 covers the follow ing parameters instrument designation chart speed programming mode chart speed limit operation timed operation scale printing time printing pen offset...

Страница 79: ...rupted The relays of the external relay mod ule ER8 remain in their current sta tus They are only operated again when the recorder is at the operating level or in the basic status Display Edit Selecti...

Страница 80: ...system designa tion The complete character set is availa ble v Section 12 5 Display Edit Selection input with keys Enter Continue with UNITWORD E TEXT _ _ _ _ _ _ _ _ _ _ Input instrument desig nation...

Страница 81: ...f a chart speed from a table of values and continuous pro gramming of the speed speed program ming in unit steps Display Edit Selection input with keys Enter Continue with KIND OF FEED PROGRAMMING E P...

Страница 82: ...es within the different chart speeds v Section 7 5 Display Edit Selection input with keys Enter Continue with FEED LIMIT E with step speed programming FEED LIMIT 7200 mm h Select chart speed 0 5 10 20...

Страница 83: ...ties within the different chart speeds v Section 7 5 Display Edit Selection input with keys Enter Continue with FEED TIME E with step speed programming FEED TIME 7200 mm h Select speed 0 5 10 20 60 12...

Страница 84: ...e set are guide values and may differ according to the chart speed Scale printing can also be started manually on the recorder Hold F key down for at least 4 sec Display Edit Selection input with keys...

Страница 85: ...printed on the chart The values which are set are guide values and differ accord ing to the chart speed Section 7 3 Display Edit Selection input with keys Enter Continue with PRINT TIME E DISTANCE 6 c...

Страница 86: ...nd from channel 2 to channel 3 This offset is compensated by a buffer storage of the measured values when the pen offset com pensation is switched on Display Edit Selection input with keys Enter Conti...

Страница 87: ...put has to operate the report v Section 8 5 10 Display Edit Selection input with keys Enter Continue with REPORT E TYPE OFF Set period of statistical report OFF DAILY WEEKLY MONTHLY EXTERN CONTACT h R...

Страница 88: ...7 6 The complete character set is availa ble v Section 12 5 Display Edit Selection input with keys Enter Continue with BEGIN TEXT E STATE ON Select status for printing text at beginning OFF ON h R E O...

Страница 89: ...ection 7 6 The complete character set is available v Section 12 5 Display Edit Selection input with keys Enter Continue with END TEXT E STATE ON Select status for printing the text at end OFF ON h R E...

Страница 90: ...the recorder The instruments are supplied fac tory set with this setting This does not apply to the custom ized linearisation tables Display Edit Selection input with keys Enter Continue with PRESETTI...

Страница 91: ...it Operation v Section 8 2 4 When customized codenumbers are input the original JUMO codenum bers are overwritten and are then no longer valid If the same numbers are defined for both codenumbers this...

Страница 92: ...al stop external speed event counter external scaling external report If a correct coden umber has been input when calling up the parameter level signal acqui sition and recording are interrupted at c...

Страница 93: ...TEXT L and TEXT H which can be freely configured report the limit transitions v Section 7 7 For lk7 and lk8 v Section 10 3 Extra Code 259 is required Accessory External relay module ER8 is required D...

Страница 94: ...EL3 SENS BR CONST Define relay position on sensor break OFF ON CONST CONST constant relay position is maintained h R E 8 8 TEXT L _ _ _ _ _ _ _ _ _ Input text for underlimit 16 characters h R l r S E...

Страница 95: ...ple change when the maths and logics module is switched on The information given in the Operat ing Manual can then become partial ly invalid Section 10 5 Display Edit Selection input with keys Enter C...

Страница 96: ...in a data network and the answer of the pen recorder Extra Code 52 or 53 is required B 70 6001 2 Display Edit Selection input with keys Enter Continue with INTERFACE E PROTOCOL J BUS Set protocol J B...

Страница 97: ...lect address 1 255 h R l r E Check 1 address 255 5 5 MIN RESPONSE TIME 000Tmsec Select minimum response time 0 999 msec h R l r E Check 0 resp time 999 h forwards R backwards Display Edit Selection in...

Страница 98: ...uired Section 7 7 Display Edit Selection input with keys Edit Continue with EXTERNAL TEXT E 2 EXTERNAL TEXT Select text number 1 8 h R E 2 2 2 CONTACT LOG INPUT 3 Select logic input OFF LOG INPUT 1 LO...

Страница 99: ...e text is printed v Section 7 7 Extra Code 259 is required Diplay Edit Selection input with keys Enter Continue with TEXT LOGIC LINKS E LINK BIN INPUT 1 4 Select inputs which are to be binary linked O...

Страница 100: ...long as the signal is present the recording is interrupted Extra Code 259 is required Display Edit Selection input with keys Enter Continue with EXTERNAL STOP E CONTACT LOG INPUT 3 Select logic input...

Страница 101: ...LOG INPUT 3 Select logic input OFF LOG INPUT 1 LOG INPUT 2 LOG INPUT 3 LOG INPUT 4 LOG INPUT 8 h R E OFF h forwards R backwards otherwise 2 2 with step speed programming FEED EXTERN 7200 mm h Select...

Страница 102: ...tered in each event counter for designation Extra Code 259 is required Section 7 8 Display Edit Selection input with keys Enter Continue with EVENT COUNTER E 2 EVENT COUNTER Select event counter 1 2 h...

Страница 103: ...from which the scale printing is initiated Section 7 4 Display Edit Selection input with keys Enter Continue with EXTERNAL SCALING E CONTACT LOG INPUT 3 Select logic input OFF LOG INPUT 1 LOG INPUT 2...

Страница 104: ...the rec ognition of the high low edge and printed out in table format In the case of a report via external contact v Section 8 5 8 the REPORT TYPE in the parameter REPORT has to be set to EXTERN CONTA...

Страница 105: ...l length 16 m Packing unit 5 rolls Part No 00331497 no name graduation linear overall length 32 m Packing unit 5 rolls Part No 00331499 no name special graduation printing as order specification overa...

Страница 106: ...9 Consumables 106 9 2 Removing and replacing the chart cassette S F ENTER E S F ENTER E...

Страница 107: ...te slot and push it up gently until the holding ejector catch snaps into position When the cassette has been inserted the paper feed should immediately take up the slack in the paper about 10mm feed I...

Страница 108: ...9 Consumables 108 9 2 1 Changing the roll chart 1 5 2 6 3 7 4 8...

Страница 109: ...9 Consumables 109 9 2 2 Changing the fanfold chart 1 4 2 5 3...

Страница 110: ...110 10 Extra Codes and accessories 10 1 Converting the chart cassette 1 3 2 4...

Страница 111: ...erface for the external relay module ER8 The 8 logic inputs can be operated through floating contacts or at the following voltage levels inactive 0 5V active 20 35V The voltage level must be applied f...

Страница 112: ...ion 10 4 The limit for each relay is programmed at the parameter level The switching differential about the limit set ting can be freely configured in the setup program The limit comparator functions...

Страница 113: ...10 Extra Codes and accessories 113 ER 8 Error Power L L L1 N 97 98 99 TxD RxD RxD TxD GND 8 1 8 2 8 3...

Страница 114: ...10 Extra Codes and accessories 114 lk7 The relay is energised when the measure ment goes above the limit setting lk8 as lk7 but relay action reversed...

Страница 115: ...232 data interchange mouse VGA graphics Software requirements Microsoft Windows1 3 1 3 11 95 NT4 0 The setup program is installed with an in stallation program for Microsoft Windows The operation of t...

Страница 116: ...nector into the recorder interrupts signal acquisition and recording In addition events are no longer covered statistical reports are aborted and reset limit values are no longer moni tored and the ou...

Страница 117: ...maximum value is record ed for as long as a specific logic input is open Closing the logic input resets the maximum value recording Thanks to the many functions operators and variables which are avai...

Страница 118: ...ce is intended for communica tion with higher level systems e g bus sys tems It is not the setup interface which is used to transfer the data between the setup program and the pen recorder The RS422 R...

Страница 119: ...chart cassette Section 9 2 1 The end of the chart has been reached Section 9 2 1 Section 9 2 2 The chart speed has been programmed as 0 mm h Section 8 2 1 The writing status is switched off OFF The s...

Страница 120: ...hing differential at the limit has been disregarded The relay status is OFF A wrong signal is being monitored against the limit No programming possible A wrong codenumber has been input The codenumber...

Страница 121: ...N 200 900 C 210 1200 C 200 600 C 270 400 C 270 1372 C 270 1000 C 270 1300 C Type Range limits Pt10Rh Pt S Pt13Rh Pt R Pt30Rh Pt6Rh B 50 1768 C 50 1768 C 0 1820 C Cold junction internal Pt100 Accuracy...

Страница 122: ...pan 0 5 mA Within these basic range limits the start and end of range can be freely programmed in 0 01mA steps Customized linearisation In the setup program there is a choice be tween linear square la...

Страница 123: ...20 240 300 360 600 720 1800 3600 7200mm h or freely programmable in 1mm h steps Chart drive by stepping motor and gearbox Chart cassette cassette for 16 m or 32 m roll chart or 16 m fanfold chart with...

Страница 124: ...ge temperature range 20 to 70 C Climatic conditions rel humidity 75 max no condensation Operating position NL 90 30 to DIN 16 257 vertical Protection to EN 60 529 Category 2 front IP54 front IP65 with...

Страница 125: ...ke up roll has not properly engaged with the chart cassette Section 9 2 1 Error or fault on recorder LOW BATTERY The battery for the real time clock and for RAM back up is discharged Please contact th...

Страница 126: ...be repeated Input a value within the range of values Date is invalid An invalid date has been input The input has to be repeated Time is invalid An invalid time has been input The input has to be rep...

Страница 127: ...to any event and can not be operated Please contact the nearest office or the factory 1 The action on probe break is configured using the setup program or the parameter relay output Section 8 6 1 Dis...

Страница 128: ...atus because key S has been pressed EXTERNAL STOP The recorder is in the Stop status because external stop has been activated by closing the appropriate logic input SETUP The setup connector has been...

Страница 129: ...0 051 3 098 b 0181 0231 052 4 099 c 0182 0232 053 5 0100 d 0183 0233 054 6 0101 e 0184 0234 055 7 0102 f 0185 0235 056 8 0103 g 0186 0236 057 9 0104 h 0187 0237 058 0105 i 0191 0238 059 0106 j 0192 02...

Страница 130: ...urements R S Number format Selection of decimal places in alpha numerical presentation R S Channel designation Designation of the channel R S Plotarea zoom Magnifier function input in R S Offset prese...

Страница 131: ...our logic inputs R S external stop A definable logic input stops the recording when a signal is applied R S external speed A definable logic input switches over to a certain chart speed R S event coun...

Страница 132: ...iguration level 3 26 92 connection 2 3 wire 4 wire circuit 17 65 67 potentiometer 17 connection diagram 16 consumables 105 current input 16 64 122 customized linearisation 64 122 cyclic display 47 D d...

Страница 133: ...tor lk7 114 lk8 114 limit mark 36 limit operation 75 limit differential of 60 112 linearisation 64 122 location 13 logic inputs 111 M maths and logics module 95 117 measurement range 65 72 measurement...

Страница 134: ...imits 60 report statistical table of statistics 87 scale printing 84 scaling 73 selection 29 service print 52 start text 88 summer time 58 text input 29 time printing 85 timed operation 83 value input...

Страница 135: ...33 limit mark 36 overwriting trace 34 presentation range offset 36 print test 45 printing mode 33 printing priority 30 recording start and end 40 relay text 42 scaling 36 scaling triggered 37 service...

Страница 136: ......

Страница 137: ......

Страница 138: ......

Страница 139: ......

Страница 140: ...36035 Fulda Germany Phone 49 661 6003 0 Fax 49 661 6003 607 E mail mail jumo net Internet www jumo net JUMO House Temple Bank Riverway Harlow Essex CM20 2DY UK Phone 44 1279 63 55 33 Fax 44 1279 63 5...

Отзывы: