5 Electrical connection
54
5.2
Assembling the signal and magnet coil cable - dual-compartment hous-
ing
Assemble both cable ends as displayed.
IMPORTANT (NOTE)!
Use ferrules.
•
Ferrules 0.75 mm
2
(AWG 19), for the shielding (1S, 2S)
•
Ferrules 0.5 mm
2
(AWG 20) for all other specifications
CAUTION – Property damage!
The shields may not touch. Otherwise, a signal short circuit occurs.
Sensor side, dimensions in mm (inch)
Transmitter side, dimensions in mm (inch)
Description
Description, wire color
Length in mm (inch)
L
1
Maximum stripped length = 105 (4,10)
(1)
Measurement potential 3, green
L
2
= 70 (2.76)
(2)
Signal line E1, violet
L
2
= 60 (2.36)
(3)
Shield 1S
L
2
= 60 (2.36)
(4)
Shield 2S
L
2
= 60 (2.36)
(5)
Signal line E2, blue
L
2
= 60 (2.36)
(6)
Data line D2, yellow
L
2
= 70 (2.76)
(7)
Data line D1, orange
L
2
= 70 (2.76)
(8)
Magnet coil M2, red
L
2
= 90 (3.54)
(9)
Magnet coil M1, brown
L
2
= 90 (3.54)
(10)
Ground wire, steel
(11)
SE clamp
(1)
(2)
(3)
(4)
(5)
(
6
)
(
7
)
(8)
(9)
(10)
(11)
(1)
(2)
(3)
(4)
(5)
(
6
)
(
7
)
(8)
(9)
(10)
(11)
Содержание flowTRANS MAG H01
Страница 2: ......
Страница 22: ...1 Safety 22...
Страница 68: ...5 Electrical connection 68...
Страница 72: ...6 Digital communication 72...
Страница 94: ...7 Startup 94...
Страница 108: ...8 Parameterization 108...
Страница 109: ...109 8 Parameterization...
Страница 110: ...8 Parameterization 110...
Страница 111: ...111 8 Parameterization...
Страница 112: ...8 Parameterization 112...
Страница 152: ...8 Parameterization 152...
Страница 160: ...9 Advanced diagnostics functions 160...
Страница 258: ...14 Annex 258...
Страница 259: ......