77
14 Operation, maintenance and care
14
Operation, maintenance and care
14.1
Cleaning
The front of the device (front foil) can be cleaned with standard detergents, rinsing and cleaning agents.
CAUTION!
The front of the device is not resistant to aggressive acids and lyes, scouring agents, and clean-
ing with a pressure cleaner.
Use of these media can cause damage.
Only clean the front of the device with suitable agents.
14.2
Sensor replacement on device versions with a separate sensor
NOTE!
Sensor replacement is not possible on JUMO digiLine electronics in device versions as a head transmit-
ter. Here, replacement of the entire module with sensor and electronics is required.
Replacement of the sensor with retention of the JUMO digiLine electronics
On device versions with a separate sensor, the sensor can be disconnected from the JUMO digiLine
electronics. If the sensor needs to be replaced, the JUMO digiLine electronics can be connected to a
new sensor and re-inserted. The "Sensor replacement function" must be used in this case to reset the
corresponding data in the JUMO digiLine electronics and increment the "sensor counter".
JUMO DSM software operating manual