75
13 Configuration
13.6
Calibration timer
Calibration timer
13.7
Display
General information
Configuration item
Selection/setting op-
tion
Explanation
Calibration interval
0 to 9999 days
Time from one calibration to the next. The time at which a cali-
bration is due is indicated by the calibration alarm on device ver-
sions with a display.
In addition, the calibration alarm from the IO-Link master can be
read and processed in the status word of the process data (see
chapter 12.1 "Process data", page 55).
Configuration item
Selection/setting op-
tion
Explanation
Language
German
English
French
Spanish
Selection of the operating language for the JUMO digiLineCR
Automatic logoff time
0 to 15 min.
The automatic logoff time can be set here. This time counts
down as soon as the user has logged in to the device. Once this
time elapses, the logged-in user is logged off automatically. If
the automatic logoff time is set to the value 0 s, then automatic
logoff is inactive. The user then remains logged in for the entire
length of the session.
chapter 7.2.1 "Log-on/Log-out", page 39
Display type
Normal
Large display
Bargraph
In the normal and large display, 2 values are shown on the dis-
play in the measuring mode (main and secondary value).
In the bargraph display, the main value is shown as a numerical
value at the center in the measuring mode and visualized un-
derneath as a bargraph display. In contrast to the normal and
large display, the secondary value is not shown in this case. The
value range of the main value for the bargraph can be set (see
next table).
The compensated conductivity is the main value and the tem-
perature is the secondary value in the default setting. However,
this setting can also be changed to meet your requirements (see
further below in this table).
Temperature Unit
°C
°F
Temperature unit setting for the device
Sign. main value
No signal
Temperature input
Compensation temp.
Uncomp. conductivity
Compens. Conductivity
Signal source for the main value display
In the measuring mode, the main value is shown on the display
as the center value (the largest).
The appearance of the display can be set in the display type
(higher up in this table).
Содержание digiLine CR HT10
Страница 2: ......
Страница 6: ...Contents...
Страница 16: ...3 Device Description 16...
Страница 26: ...4 Identifying the device version 26...
Страница 50: ...10 Troubleshooting in the event of faults 50...
Страница 70: ...12 IO Link data 70...
Страница 78: ...14 Operation maintenance and care 78...
Страница 83: ......