14 Calibrating Redox sensors
186
14.3.2
Two-point Calibration
Step
Action
1
Ensure that
•
the calibration default settings are correct
•
"Percent" is set as the Redox unit in the configuration of the Re-
dox measuring input.
v
Chapter 14.2.2 "Calibration default settings for Redox sensors",
v
Chapter 10.5.3 "pH/Redox/NH3 analysis inputs", Page 126
2
Start the two-point calibration.
Device Menu
r
Calibration
r
Select Analysis Input for pH/Redox/
NH
3
r
Two-point calibration
3
Enter the concentration value of the first reference solution in per-
cent. Confirm by pressing "OK".
4
Clean the Redox electrode and immerse it in the first test solution.
Wait until the measuring value displayed stabilizes and then con-
firm the result of the measurement by pressing "OK".
5
As in step 3, enter the concentration value of the second solution
in percent. Confirm by pressing "OK".
6
Clean the Redox electrode and immerse it in the second
test solution. Wait until the measuring value displayed stabilizes
and then confirm the result of the measurement by pressing "OK".
7
A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
8
Press "Yes" to accept the calibration values determined and enter
the calibration in the calibration logbook.
Press "No" to discard the results.
Tap button to enter
field
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......