14 Calibrating
Redox sensors
14.1
Notes
14.2
General information
The calibration of Redox sensors is based on measurements in test solutions
with a defined Redox potential.
14.2.1
Calibration methods for Redox sensors
Zero-point calibration
This calibration method is used to determine the Redox zero point.
A test solution with a defined Redox potential is needed as a reference.
"mV" must be set as the Redox unit in the configuration of the Redox measuring
input.
v
Chapter 10.5.3 "pH/Redox/NH3 analysis inputs", Page 126
Two-point Calibration
This calibration is used to establish an application-specific measurement char-
acteristic curve where Redox potentials are represented as a percentage of con-
centration values. The Redox potentials of 2 solutions are measured.
Concentration values in percent are then assigned to the measured values by
the user.
Two process-typical sample solutions are needed as calibration solutions as a
reference.
"Percent" must be set as the Redox unit in the configuration of the Redox mea-
suring input.
v
Chapter 10.5.3 "pH/Redox/NH3 analysis inputs", Page 126
Example:
In a detoxification plant, the toxicity of a liquid is to be measured on
the basis of the Redox potential. The calibration requires 2 solutions:
•
the concentration of the highly toxic solution is given as 80 % by the user, for
instance
•
the concentration of the detoxified solution is given as 10 % by the user, for
instance
Using the Redox potential, the toxicity can now be measured and displayed in
percent.
WARNING!
During the calibration, the relays and analog output signals assume the config-
ured states!
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......