13 Calibrating a pH measuring chain
178
13.3
pH Calibration routines
13.3.1
Zero-point calibration
NOTE!
You must be logged in with corresponding user rights to perform calibrations.
v
Chapter 8.2.1 "Log-on/Log-out", Page 78
Step
Action
1
Start the Zero-point calibration.
Device Menu
r
Calibration
r
Select analysis input for pH/Redox/
NH
3
r Open Zero-point calibration
2
If temperature compensation was not specified in the calibration
default settings, enter the temperature of the Buffer solution here
manually.
If temperature compensation was specified, the
temperature of the Buffer solution is determined automatically.
3
Clean the pH electrode and immerse it in the Buffer solution.
tap button to enter
temperature
Содержание AQUIS touch P 202580
Страница 12: ...Content...
Страница 26: ...4 Identifying the device version 26...
Страница 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Страница 32: ...5 Mounting 32...
Страница 64: ...7 Startup 64...
Страница 100: ...1 2 3 4 5...
Страница 107: ...107 8 Operation Example of a Quad View...
Страница 108: ...8 Operation 108...
Страница 114: ...9 Parameterization 114...
Страница 160: ...10 Configuration 160...
Страница 182: ...13 Calibrating a pH measuring chain 182...
Страница 196: ...16 Calibrating CR conductivity sensors 196...
Страница 206: ...17 Calibrating Ci conductivity sensors 206...
Страница 214: ...18 Calibrating universal inputs 214...
Страница 228: ...19 PC Setup Program 228...
Страница 319: ...319 20 Technical data...
Страница 320: ...20 Technical data 320...
Страница 321: ......