81
8 Calibrating sensors
8.3.3
Two-point calibration
1
Prerequisite
• The device is supplied with voltage,
see chapter 5 "Installation", page 17 et seqq.
• The sensor is connected and has completed a run-in time (2 hours for sen-
sor type 202630/53)
• Sensor startup on the device is completed
• Calibration has been enabled, see chapter 6.7 "Administrator level", page
• the device is in "Measurement mode".
Selecting calibration methods
✱
Initiate calibration using the
key (or via the administrator level).
✱
Select "2-POINT CALIB." and start calibration with the
key.
The device display changes to measurement of the zero point.
1
Using the example of the sensor for free chlorine, type 202630/53
CAUTION!
The response of the outputs during calibration depends on their configuration.
Содержание AQUIS 500 RS