53
17 Calibrating u
niversal inputs
17.1
Important information
17.2
General information
17.2.1
Calibration methods for universal inputs
Universal inputs can be configured with various operation modes for a number
of different process variables (see table below).
Detailed information on the possible configurations:
v
Chapter 9.5.2 "Universal inputs of base unit and optional boards", Page 99
Appropriate calibration routines for each operation mode of a universal input can
be enabled in the calibration default settings.
The following table lists the availability of calibration routines for the individual
operation modes.
Zero-point calibration
This calibration method is used to determine the pH zero point on a measure-
ment characteristic curve. The slope is retained.
A test solution with a defined value of the respective process variable is needed
as a reference.
Slope calibration
This calibration method is used to determine the slope of a measurement char-
acteristic curve. The zero point is retained.
A test solution with a defined value of the respective process variable is needed
as a reference.
WARNING!
During the calibration, the relays and analog output signals assume the config-
ured states!
linear
scaling
Temp. mea-
surement
pH value
measure-
ment
Conductivity
measure-
ment
free chlorine,
ph/T-com-
pensated
Zero-point calibration
X
X
Slope calibration
X
X
Two-point calibration
X
X
Rel. cell constant
X
Temperature coefficient
X
TC-curve
X
Universal input
Operation modes
Calibration
routine
Содержание 202580
Страница 6: ...Content...
Страница 28: ...12 Calibrating a pH measuring chain 28...
Страница 42: ...15 Calibrating CR conductivity sensors 42...
Страница 52: ...16 Calibrating Ci conductivity sensors 52...
Страница 60: ...17 Calibrating universal inputs 60...
Страница 79: ...79 18 Technical data...
Страница 80: ...18 Technical data 80...
Страница 81: ......