26 Configuration level (not instrument-specific)
78
26.6 Other outputs I - C213
*The factory-set parameters are shown in the position boxes.
a
b
c
d
C213*
8
0
3
0
Function of output 3
(relay 3 or analog output)
I
I
I
I
No function
0
I
I
I
Hold (relay
only)
1
I
I
I
Alarm pulse contact
(relay only)
2
I
I
I
Alarm steady contact
(relay only)
3
I
I
I
MAX temperature limit comparator
(relay only)
4
I
I
I
MIN temperature limit comparator
(relay only)
5
I
I
I
MAX pH / redox limit comparator
(relay only)
6
I
I
I
MIN pH / redox limit comparator
(relay only)
7
I
I
I
Process value pH
(analog output only)
8
I
I
I
Process value temperature
(analog output only)
9
I
I
I
Proportional controller 1
(analog output only)
1
A
I
I
I
Proportional controller 2
(analog output only)
1
b
I
I
I
I
I
I
Signal for output 3
(analog process value output only)
2
I
I
I
0 — 20 mA
0
I
I
4 — 20 mA
1
I
I
0 — 10 V
2
I
I
2 — 10 V
3
I
I
20 — 0 mA
4
I
I
20 — 4 mA
5
I
I
10 — 0 V
6
I
I
10 — 2 V
7
I
I
I
I
Function of output 4 (logic output)
I
I
No function
0
I
Hold
1
I
Alarm pulse contact
2
I
Alarm steady contact
3
I
MAX temperature limit comparator
4
I
MIN temperature limit comparator
5
I
MAX pH / redox limit comparator
6
I
MIN pH / redox limit comparator
7
I
I
Alarm monitoring of relays K1 and K2
3
I
K1
/
K2
I
monitored
monitored
0
monitored
not monitored
1
not monitored
monitored
2
not monitored
not monitored
3