Maintenance
24
9 Maintenance
9.1 Replace detachable power cord
The device is equipped with a detachable power cord.
If the power cord needs to be replaced, ensure that the new one is at least
dimensioned for the device power requirements. Insufficiently dimensioned
power cords must not be used. See type plate for mains voltage and current
value.
We recommend only using original JULABO spare parts.
9.2 Check safety symbols
The safety labels affixed to the device must be clearly legible at all times. Their
condition must be checked every two years.
1. Check the safety signs on the device for legibility and completeness.
2. Replace defective or missing safety markings.
Safety signs can be reordered from JULABO.
The safety signs on the device have been checked.
9.3 Clean device
Clean the device regularly to ensure that it operates without issue over the long-
term.
Lint-free cloth
Mild cleaning agent
Limescale cleaner
NOTE
Damage to the electronics due to water penetration!
Ingress of water can damage electronic components of the device and thus
lead to failure of the device.
• Clean the outside of the device with a damp cloth only
• Prevent water from entering the device