![Julabo Magio MX Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/julabo/magio-mx/magio-mx_original-operating-manual_2030737057.webp)
Operation
57
Factory setting for the administrator password: 000000.
4. Press the
<Reset>
button.
5. Acknowledge the confirmation prompt with
[OK]
.
All unit account settings are reset to factory settings.
8.5 Record data
8.5.1 Record measurement data
Measurement data for an ongoing temperature control application can also be
recorded onto a USB stick at the same time. The record documents the date,
time, setpoint temperature, intern actual value, external actual value and the
percentage power. The data is saved as a .txt file and can be subsequently
analyzed.
The device is switched on.
A USB stick is available.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Record data>
.
The warning appears: “No USB stick recognized!”
3. Insert the USB stick into the USB socket on the back of the device.
4. Press the button
[File]
to change the filename if required.
5. Press the button
[Sampling time]
and set the cycle in which measurement
data should be recorded.
6. Activate the
[Active]
button.
The measurement data recording is started and this is shown with a red
point on the home screen. The data is recorded on the USB stick using the
specified filename. Stop recording by deactivating the
[Active]
button.
8.5.2 Reading out blackbox data
The blackbox stores all relevant data from the last 30 minutes. In addition, the
black box logs alarms and warnings.
The blackbox can be read out and the data sent to technical service for analysis.
The device is switched on.
A USB stick is available.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Record data>
.
The warning appears: “No USB stick recognized!”
3. Insert the USB stick into the USB socket on the back of the device.