Operation
56
8.4.6 Lock the device
A logged-in user can lock the device to prevent unwanted interventions. Active
temperature control application is not interrupted. A locked device cannot be
operated, it must first be unlocked by the last user or an administrator.
The device is switched on.
A basic user or advanced user is logged in.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Unit account>
.
The logged in user group is displayed.
3. Activate the
[Lock device]
button.
The device is unlocked. The lock symbol is displayed at the bottom edge of
the home screen.
8.4.7 Set inactivity timer
After a defined period of inactivity on the display, the device is automatically
locked. The inactivity timer can be set up by a logged-in user.
The device is switched on.
A basic user or advanced user is logged in.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Unit account>
.
The logged in user group is displayed.
3. Press the
[Inactivity duration]
button.
An input field will open.
4. Enter the duration for the inactivity timer in the specified time format and
confirm your input.
The duration is displayed on the
[Inactivity duration]
button.
5. Activate the
[Inactivity timer]
button.
The inactivity timer is set up and active. As soon as the duration of inactivity
on the display has elapsed, the device is locked.
8.4.8 Reset settings
All settings in the
<Unit account>
menu can be reset to factory settings at any
time. The administrator password is required to reset the settings.
The device is switched on.
1. Call up the
<Main menu>
.
2. In the submenu
<Unit account>
call up the menu option
<Settings>
.
An input field will open.
3. Enter the administrator password and confirm your input.