![Julabo MA-12 Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/julabo/ma-12/ma-12_operating-manual_2030709040.webp)
Menu functions
9.6. MENU LIMITS
Menu level 1
>
Li
< = LIMITS – limitation of operating temperature range
Maximum and minimum setpoint
Restriction of the adjustable temperature range
.
Level 2
Parameter level
Press the
key if parameter is to be retained.
Correction function for parameters or values (prior
to OK).
•
Integer digits flash, set by pressing
+
•
Decimal digits flash, set by pressing
+
•
Integer digits flash, set by pressing
+
•
Decimal digits flash, set by pressing
+
Factory settings::
200 °C
(Setpiont High)
-50 °C
(Setpoint Low)
The limitation of the operating temperature range effects the
temperature setting in the menu with the key
.
Only setting of working temperatures which lie within the
determined limits is possible
This applies to settings in the MENU
,
(refer to page 22)
and for settings in the MENU
high temperature low temperature
(refer to page 27)
The temperature values are automatically deferred into the limit
range.
Setting range: -50,0 °C ... +200,0 °C
40