Appendix
89
Auxiliary parameters
System response
in_hil_00
Set actuating variable limit for cooling capacity (%)
in_hil_01
Set actuating variable limit for heating capacity (%)
11.1.2 OUT commands
OUT commands set device parameters. Remote control mode must be active.
Parameter
settings
Parameter
Setting
out_sp_00
xxx.xx
Setting for the setpoint temperature
out_sp_03
xxx.xx
Setting for the high temperature
out_sp_04
xxx.xx
Setting for the low temperature
out_sp_06
xxx.xx
Setpoint temperature setting via watchdog function
out_sp_07*
x
Set pump to preset levels 1…5
out_sp_10
xxx.xx
Setting for the actuating variable via the serial
interface
out_sp_11
x
Setting the temperature units:
0 = °C
1 = °F
out_sp_27
xxx
Pump capacity in %
out_sp_28
xxx.xx
High temperature alarm limit setting
out_sp_29
xxx.xx
Low temperature alarm limit setting
* Do not use for reprogramming, use out_sp_27 instead. Pump stage 4 has an output pressure of
approx. 450 mbar.
Device modes
Parameter
Setting
out_mode_01
x
Setting 0 (for reasons of downwards compatibility
the parameters 1 or 2 can also be used. Within the
device it has no effect)
out_mode_03
x
Setting of the operating mode EPROG:
0 = 0-10 V
1 = 0-20 mA