Operation
47
3. Use the arrow keys to set the value after the decimal point and confirm with
[OK]
.
The set value is applied. The new setpoint temperature flashes briefly.
The setpoint temperature is set and active. Setpoint temperature
configuration can be interrupted using the
[MENU]
key. The previous value
is then retained.
8.4 Activate autostart function
The autostart function makes it possible to start a temperature control
application directly using the mains switch or via an intermediate timer.
The device is configured ex works in such a ways that it switches to a safe
operating status in the event of power failure. The autostart function is
deactivated. The display shows “OFF.” The refrigeration aggregate, heater, and
pump motor are disconnected from the mains voltage.
The device is switched on.
1. Press the
[MENU]
key.
2. Use the arrow keys to select the menu option
<conF>
and confirm with
[OK]
.
3. Use the arrow keys to select the menu option
<Auto>
and confirm with
[OK]
.
4. Use the arrow keys to select
<On>
and confirm with
[OK]
.
The display flashes briefly.
The autostart function is activated. The next time the device is switched on,
the temperature control application will immediately start with the preset
values. You can also insert and program a timer. In this case the mains
switch of the device must remain on.