![Julabo CORIO CP-310F Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/julabo/corio-cp-310f/corio-cp-310f_original-operating-manual_2030701020.webp)
Product description
20
5.4.1 Material of parts that come into contact with the medium
The table lists parts that could come into contact with the bath fluid as well as
the material that the parts are made of. This data can be used to check the
compatibility of the parts with the bath fluid used.
Parts that come into contact with the
medium
Material
Bath tank
1.4301/304H
Bath tank drain opening
1.4301/304H
Bath tank/bath cover gasket
FKM Viton
O-ring drain valve
FKM Viton
5.4.2 Refrigerant
For safety reasons and in case of leakage in the refrigerant cycle, there is a
specified room volume per kg of refrigerant permitted at the installation site so
that no flammable refrigerant/air mixture can form. The amount of refrigerant is
indicated on the nameplate.
1 m³ of space is required for 0,008 kg of R290 refrigerant.
1 m³ of space is required for 0.357 kg of R449A refrigerant.
1 m³ of space is required for 0.423 kg of R452A refrigerant.