8.4 Start temperature control application
You can start a temperature control application right on the device, or program
one using the timer.
The unit is ready for use.
1. Switch the unit on at the mains switch.
2. Use the arrow keys to set the desired setpoint temperature.
3. Press and hold the
[OK]
key until the temperature control application starts.
The setpoint temperature is saved. The display flashes briefly. The unit starts
the temperature control application at once. The temperature control
application can be stopped with the
[OK]
key.
Observe the following for heating circulators:
For temperature control applications near or below the ambient
temperature: Use a cooling coil or JULABO immersion cooler.
8.5 Activate autostart function
The autostart function makes it possible to start a temperature control
application directly using the mains switch or via an intermediate timer.
The device is configured ex works in such a ways that it switches to a safe
operating status in the event of power failure. The autostart function is
deactivated. The display shows “OFF.” The refrigeration aggregate, heater, and
pump motor are disconnected from the mains voltage.
The unit is switched off.
The autostart function is deactivated.
1. Simultaneously press and hold the
[OK]
key and the power switch until the
device is switched on.
The display shows
<AOn>
.
The autostart function is activated. The temperature control application
starts immediately with the preset values, each time the device is switched
on, as long as the autostart function is active. To deactivate the autostart
function, switch off the device and repeat the procedure. The display will
then show
<AOFF>
.
You can also insert and program a timer. In this case the mains switch of the
device must remain on.