8.7. Evaluation of the temperature limits
The high- and low-temperature limit can be adjusted in two ways
(see page 66)
3. As pure warning function with an acoustic warning signal in
regular intervals.
Adjustment „Li 0“ – factory adjustment
4. As temperature limit by switching-off the heating. Adustment „Li
1“
The alarm is indicated by optical and audible signals
(continuous tone) and on the MULTI-DISPLAY (LED) appears
the error message "
E
rror
01
".
1.
Press
and
at the same time.
2.
Use the cursor keys
to select the menu option
Limit. (example: „Li 0“).
3.
Select the alternative state with the edit keys
and
confirm the selection with the ENTER key
(display „Li 1“).
4.
Press
and
at the same time.
9.
Safety installation (with shutdown function)
))))
(excess temperature protection)
These safety installations is independent of the control circuit.
When the temperature of the bath liquid has reached the safety
temperature, a complete shutdown of the heater and pump (only
CW-models) is effected.
The alarm is indicated by optical and audible signals (continuous
tone) and on the MULTI-DISPLAY (LED) appears the error
message "
E
rror
01
".
Содержание BAT7146
Страница 12: ...Verantwortung des Betreibers Sicherheitshinweise 10 2 2 EG Konformit t...
Страница 54: ...Operator responsibility Safety instructions 52 2 2 EC Conformity...
Страница 86: ...Cleaning repairing the unit 84...
Страница 96: ...Responsabilit de l utilisateur Informations de s curit 94 2 2 C E E Conformit...
Страница 128: ...Nettoyage de l appareil Entretien 126...
Страница 138: ...Responsabilidad del propietario aspectos relacionados con la seguridad 136 2 2 Conformidad CE...
Страница 170: ...Limpieza reparaci n del aparato 168...