6. Maintenance and Troubleshooting
The iAED-S1 is very simple to maintain. The defibrillator performs a self-test
every day and its extensive automatic self-test features eliminate the need for
any manual calibration. The iAED-S1 has no user-serviceable parts.
6.1 Routine Maintenance
To ensure that the device is ready for use at any time, routine maintenance
must be performed by users, who should fill in the periodic inspection table
(see Annex H.). Table 2 shows the recommend routine maintenance plan.
Table 2 Periods of routine maintenance
Check Items
Daily
Monthly
After Each Use
Checking the status indicator
●
●
●
Checking the expired date of the pads
●
Checking the Completeness of Parts
●
●
6.1.1 Checking the Status Indicator
Procedure
Automatically perform daily/weekly/monthly self-test.
Verification
⚫
The status indicator is green: ready for use.
⚫
The status indicator is red: abnormal. Do the followings:
Press
(the
information button) to get detailed information:
◼
If it prompts “Battery low”, replace the battery immediately.
◼
If it prompts “Service required”, contact Jousing Medical for service.