8.4.5 SET
CALIBRATION
OFFSET
This function allows the user to adjust offsets in the calibration level when the tank is full
of water. Normally the test water level reading with the tank full of water is 4.0 inches. If
the test water level is not 4.0 inches use this function to adjust the test water level
reading to 4.0 inches. The maximum offset is +/- 2.5 inches of water. Normally after a full
calibration this calibration offset will be set to 0.0 inches.
CAL MODE–USE UP/DWN
KEYS TO SET CAL
OFFSET, ENTER TO
FINISH. +X.X INCHES
Pressing the Enter button will display:
CAL MODE–CAL OFFSET
INCHES ENTERED
SET TO: +X.X INCHES
The unit returns after a few seconds to automatic or manual display modes.
8.4.6 CLEAN CYCLE LIMIT
This function allows the user to override the automatic clean cycle limit of no more than
one clean cycle every 12 hours. After each clean cycle the clean cycle limit is enabled.
Note:
Disabling this function is typically only used for test purposes.
CAL MODE–USE UP/DWN
KEYS TO ENABLE OR
DISBALE CLEAN CYCLE
12HR LIMIT–DISBALED
Pressing the Enter button will display:
CAL MODE–CLEAN CYCLE
12 HR LIMIT SET TO
-DISABLED
GI-2000 Manual V2.3
Page
33
The unit returns after a few seconds to automatic or manual display modes