32 DePuy Synthes User Manual Small Battery Drive II
Care and Maintenance
Cleaning and Disinfection
Warning:
The Small Battery Drive II batteries
must not be cleaned following the Manual
Cleaning Instructions.
1.
Remove debris.
Rinse the device under running
old tap water for a minimum of 2 minutes. Use a
soft lint-free cloth or soft-bristled brush to assist in
removing gross soil. For cannulations of the
handpiece and attachments, the cleaning brush
(519.40) shown below should be used.
Notes:
•
Brushes and other cleaning tools shall be either
single use items or, if reusable, be decontaminated
at least daily using a solution as detailed in section
“3. Spray and wipe”.
•
Brushes shall be inspected before daily use and
discarded if they have degraded to the point where
they may scratch instrument surfaces or be in-
effective due to worn or missing bristles.
Warnings:
•
Do not immerse the handpiece, batteries, battery
casings or attachments in aqueous solutions or in
an ultrasonic bath.
•
Do not use pressurized water as this will cause
damage to the system.
Precaution:
Do not use pointed objects for cleaning.
2.
Manipulate moving parts.
Manipulate all
moving parts such as the triggers, sleeves
and switches under running tap water to
loosen and remove gross debris.
Cleaning brush (519.40)
Manual Cleaning Instructions