Surgery Start Up
Z370584 Rev. D
End Case
4-23
Figure 4-21 — End Case Screen
NOTE: For ease of viewing, the EPT,
EFX, UST, and DT times are in a large font size on the
End Case screen. The EFX shows only when you attach an ELLIPS FX or VERITAS
Swivel handpiece.
The EPT or EFX timer appears on the upper left side of the surgical screen. You can tap the
timer to view the detailed timer pop-up with EPT, EFX, UST and DT timers of the surgical
case. Each timer indicates the following:
•
Effective Phaco Time (EPT). EPT is ultrasound time as a weighted total that takes into
account the amount of power being used:
at 100% power: 1 sec. U/S Time = 1 sec. EPT
at 50% power: 1 sec. U/S Time = 0.5 sec. EPT
Average phaco power (AVG) = EPT/UST
•
Ultrasonic (U/S) time (UST) in foot pedal position 3
•
EFX when you attach an ELLIPS FX or VERITAS Swivel handpiece
•
DT – Diathermy Time
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...