Prime/Tune
Z370584 Rev. D
4-14
Surgery Start Up
NOTE: Ensure you have filled the tubing with fluid using the Continuous Irrigation
feature. If the tubing does not have fluid, the amount dispensed during Cup Fill may not
be the amount selected.
4. Tap
Start
to turn on Cup Fill.
5. You can tap
Stop
to turn off Cup Fill. This stops the process before the system dispenses
the requested amount of fluid.
6. Perform a phaco prime/tune. The Prime/Tune screen indicates the progress of the prime/
tune process.
Prime/Tune
You must prime/tune:
•
before each procedure
•
anytime you reattach the phaco handpiece
•
after you have inserted or replaced a tubing pack
The prime/tune process fills the I/A tubing with fluid, performs a vacuum check, and tests
and characterizes the phaco handpiece.
You can select
Continuous Irrigation
to allow fluid to free flow from the bottle to collect
fluid.
The design of I/A Prime allows for a procedure that does not require a phaco handpiece. The
design of tune allows for a quick tune of the phaco handpiece with an I/A tubing set. You can
use tune if you replace a phaco tip during a procedure.
You can shorten the overall prime sequence by selecting
Bypass
if you used Continuous
Irrigation or primed the I/A tubing.
To access the prime/tune routines, tap the
Prime/Tune
button. The console shows the
Prime/Tune screen with all the prime and tune options.
NOTE: Prime/Tune is a combination of tune followed by an I/A prime.
WARNING!
Do not have the handpiece tip in the eye of the patient when you prime and tune the
handpiece.
CAUTION!
If you do not properly prime the I/A tubing, errors can occur.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...