Warnings, Cautions, and Other Residual Risks
Z370584 Rev. D
Warnings
2-5
•
This unit complies with all Electromagnetic Interference (EMI) standards and
requirements. It is possible that interference provided by the operation of the HIGH
FREQUENCY (HF) SURGICAL EQUIPMENT can adversely influence the operation of other
electronic equipment.
•
Do not have skin-to-skin contact on the patient. For example, between the arms and the
torso. Insert dry gauze to avoid contact, as appropriate.
The unit does not contain any neutral electrode.
The diathermy output is bipolar.
JJSV recommends that you check the condition of all interconnecting and handpiece
cables on a regular basis.
•
Risk of burns and fire. Do not use the system near conductive materials such as metal bed
parts, inner spring mattresses, or similar items. Replace electrode cables on evidence of
deterioration.
•
Hazardous electrical output. This equipment is for use only by qualified personnel.
•
If there is damage to the cord or the plug, do not use the instrument. A damaged cable can
cause an electric shock to the user or a fire hazard to the system. Call JJSV customer
service to order a new cord.
•
Disconnect the power before you service the equipment.
•
Remove the power cord from the power outlet when the equipment is not in use.
•
Do not obstruct the power outlet so you can readily remove the power cord.
•
To prevent the risk of fire or damage to the instrument, replace the fuses with the exact
type and rating (check the voltage sticker on the back panel of the system to confirm your
system voltage).
•
Not recommended for use in condensing environments. If exposed to a condensing
environment, allow the system to equilibrate to typical operating room conditions prior to
use. Only use the system within the stated relative humidity conditions.
•
You do not need to use a NEUTRAL ELECTRODE with this HIGH FREQUENCY (HF)
SURGICAL EQUIPMENT.
•
Failure of the HIGH FREQUENCY (HF) SURGICAL EQUIPMENT could result in an
unintended increase of output power.
•
Do not replace the foot pedal battery while in use with a patient.
•
Do not replace the remote control battery while in use with a patient.
•
Do not try to replace the batteries for the Advanced Control Pedal. Call your JJSV
technical service representative to replace the batteries.
•
Sterility assurance is the responsibility of the user. You must sterilize all non-sterile
accessories prior to use.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...