Status, Warning, and Error Messages
Z370584 Rev. D
10-10
Error Messages, Troubleshooting, and Diagnostics
No sound when using diathermy
1.
Ensure the volume setting is at a level of 6 or greater on Sounds screen.
2. Check for sounds when you push any touchscreen or remote buttons.
3. Check for an audible confirmation upon completion of system start-up test (at power up).
Vitrectomy
No vitrectomy cutting or poor cutting
1.
Ensure you selected the Vitrectomy mode on the touchscreen.
2. Verify that the surgeon is in foot pedal position 3, if using IAC foot pedal mode for
vitrectomy. If using ICA, verify the foot pedal is in position 2.
3. Check the foot pedal operation.
4. Check the tubing connections to the vitrectomy cutter.
5. Check the vitrectomy tubing connection to the front panel receptacle on the system.
6. Check the vitrectomy rate (cpm) setting on the touchscreen. Lower the cpm, if
necessary.
7. Check that irrigation and aspiration are working correctly.
8. Verify that the cutter blade moves.
9. Replace the vitrectomy cutter and try again.
Status, Warning, and Error Messages
The system shows status, warning, and error messages on the monitor.
Alerts (Gray) show at the top of the screen. You do not need to clear an alert as with an error
message. An alert, for example, can be: Not Primed/Not Tuned.
Error messages show at the top of the screen. Tap
to open the corrective action for that
error. After you correct the error, tap
to clear the error message from the screen.
The message can show possible solutions to the error or recommendations to clear the error.
The messages can indicate the available options if a component or subsystem fails.
The following list of error messages have corrective actions that you can take to clear the
error.
Error Messages
101 Fluidics communication error.
Probable cause: Invalid data or communication error.
Corrective action:
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...