Program Settings
Z370584 Rev. D
FUSION Fluidics Phaco
6-35
CASE One Touch
To simplify the programming of the CASE function, you only need to define the basic CASE
parameters once. You can adjust the CASE function quickly and easily from the CASE One
Touch settings on the surgical screen. Use these controls to change the CASE functionality
for greater efficiency (up arrow) or more control (down arrow) to suit any particular
combination of cataract density, surgical technique, or personal preference.
When CASE is On, use the
One Touch
buttons to adjust the CASE parameters.
Figure 6-24 — CASE One Touch
Table 6.8 – CASE One Touch Parameter Settings
Reflux
Reflux is the controlled back flow of fluid through the aspiration port of the handpiece.
Reflux is used to gently release or dislodge unwanted material from the handpiece tip. Reflux
can also be used to “tent” the incision site to allow easier tip insertion. Reflux is not intended
to clear a clogged handpiece. However, reflux can be used to identify a blockage.
The reflux action can be programmed on any available foot pedal switch. This causes fluid to
be expelled from the aspiration line into or towards the eye.
The reflux is active until the foot pedal switch is released.
Parameter
CASE -2
CASE -1
CASE STD
CASE +1
CASE +2
Pump Ramp Setting
Program
default
Program
default
Program
default
CASE STD
+10%
CASE STD
+20%
CASE Occlusion Delay
CASE STD
-200 ms
CASE STD
-100 ms
Program
default
CASE STD
+100 ms
CASE STD
+200 ms
CASE Upper Threshold
CASE STD
-5%
Program
default
Program
default
Program
default
CASE STD
+5%
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...