FUSION Fluidics Phaco
Z370584 Rev. D
6-30
Program Settings
Figure 6-20 — FUSION Settings Screen
7. Tap
Finished
to close the screen.
8. Tap
Save
to retain the settings.
9. Tap
Exit
to close the screen.
Occlusion Vacuum Threshold
In Occlusion mode phaco, you can set an occlusion threshold value for vacuum.
When in Occlusion mode phaco, there is an additional control panel for vacuum. The vacuum
threshold setting lets you choose the vacuum level at which occluded settings take effect.
To adjust the occluded vacuum threshold, tap the up or down arrows. When you decrease
the threshold, the occlusion settings take effect sooner.
Occlusion Aspiration Rate
In Occlusion mode phaco, you can set a different maximum flow value for aspiration. There is
an additional control panel for aspiration below the standard aspiration control panel.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...