Default Settings
Z370584 Rev. D
6-18
Program Settings
Table 6.6 – 21 Gauge Program – IA, Vitrectomy, and Diathermy Settings
Max power
Unoccluded (%)
50
40
35
20
Power delivery
Continuous
Continuous
Continuous
Continuous
WHITESTAR setting (duty cycle)
On 6/3 (67%)
On 8/4 (67%)
On VWS (67%)
Quadrant 1 - 12/6
Quadrant 2 - 10/5
Quadrant 3 - 8/4
Quadrant 4 - 6/3
On 4/8 (33%)
Occluded mode (%)
50
40
40
20
Power delivery
Continuous
Multi-Burst
Multi-Burst
Continuous
WHITESTAR setting (duty cycle)
On 6/3 (67%)
On 8/4 (67%)
On 8/4 (67%)
On 4/8 (33%)
Power type, occluded and
unoccluded
Linear
Unoccluded - Linear
Occluded - Panel
Unoccluded - Linear
Occluded - Panel
Linear
Short pulse rate, occluded and
unoccluded
1–14 pps
6 pps
6 pps
6 pps
6 pps
Long pulse rate, occluded and
unoccluded
1–6 pps
4 pps
4 pps
4 pps
4 pps
Pulse Shape Setting (WHITESTAR
ICE)
On, Off
Off
Off
Off
Off
Pulse Shape Parameters
% Kick Low
5
5
5
5
% Kick High
5
5
5
5
Low Power Limit
0
0
0
0
High Power Limit
80
80
80
80
Initial Mode
On, Off
Phaco (first enabled
submode)
Surgical
Mode/
Submode
Aspiration Rate cc/
mm (Min/Max) for
Peristaltic Pump type
Vacuum mmHg (Min/
Max), Pump Type
Power % / Cut Rate
cpm (Min/Max)
WHITESTAR
(ICE) Pulse
Shaping
Pump/
Vac Ramp
IV Pole
Height
(cm)
IA1
36 Panel
0 / 600
Linear, Venturi
NA
NA
60% / 3
85
IA2
5 Panel
0 / 10
Linear, Peristaltic
NA
NA
20% / 1
85
IA3
50 Panel
0 / 600
Linear, Venturi
NA
NA
90% / 4
95
ICA
(Vitrectomy)
20 Panel
300 Panel, Venturi
50 / 2000 cpm
Panel
NA
75% / 3
50
IAC
(Vitrectomy)
20 Panel
0 / 350
Linear, Venturi
50 / 500 cpm
Panel
NA
75% / 3
60
DIA 1
NA
NA
5 / 30
Linear
NA
NA
NA
DIA 2
NA
NA
5 / 30
Burst
NA
NA
NA
Parameter
Available
Settings
Default Settings
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...