Foot Pedal Setup
Z370584 Rev. D
5-30
Surgeon Settings
Advanced Control Pedal Calibration and Testing
Follow the steps below to calibrate and test your foot pedal.
NOTE: In addition to the procedure below, you can access the Advanced Control Pedal
Summary screen by pressing the foot pedal icon (shown below) near the top of the
screen.
1.
On the Surgeons and Programs screen, tap
.
2. Tap the
Calibration
button.
Figure 5-25 — Advanced Control Pedal Calibration
3. With your foot off the foot pedal, tap the
Phase 1
button.
4. Put your foot on the foot pedal, press and hold the foot pedal all the way down and tap
the
Phase 2
button.
5. Move the pedal to the far left and hold that position while you tap the
Phase 3
button.
6. Move the pedal to the far right and hold that position while you tap the
Phase 4
button.
Содержание Veritas
Страница 1: ...Z370584 Rev D 09 2020 Johnson Johnson Surgical Vision Inc 2020 Software Release 8 000 OPERATOR MANUAL...
Страница 2: ...Z370584 Rev D This page was intentionally left blank...
Страница 4: ...Z370584 Rev D This page was intentionally left blank...
Страница 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Страница 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Страница 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Страница 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Страница 124: ...Create a New Program Z370584 Rev D 6 2 Program Settings Figure 6 1 Surgeons and Programs Screen 1 On Select Program tap...
Страница 146: ...WHITESTAR Technology Z370584 Rev D 6 24 Program Settings Figure 6 14 Duty Cycles for Variable WS...
Страница 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Страница 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Страница 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Страница 270: ...Z370584 Rev D I 8 This page was intentionally left blank...