PAGE 87 of 185
CX0540 Rev 00.1
3.
Loosen the probe holder adjustment knob and remove the outer probe
holder arm
(Fig. 220)
.
4.
Loosen the arm clamp screw
(Fig. 221).
5.
Place the wedge on the pivot button of the inner probe holder
arm
(Fig. 221).
6.
Align the middle of the wedge with the centre of the yoke
(Fig. 222).
7.
Tighten both the probe holder adjustment knob and the arm clamp screw
(Fig. 222)
while ensuring the wedge remains centred with the yoke.
Fig. 220 - Remove outer arm
Fig. 221 - Adjust inner arm
Fig. 222 - Align probe with yoke
Fig. 223 - Tighten knob and screw