To configure the NAVIC system for scanning using a single module and a 3-axis
nozzle scanning system, follow these steps
(see “3-Axis Nozzle Scanning” on page
91)
:
1.
Separate the crawler’s drive modules
(see “Connecting/Disconnecting Left and
Right Modules” on page 71)
.
2.
Mount and connect the
5
umbilical at a 90° angle
to
3
right drive module.
3.
Connect the
5
umbilical
to the
9
power controller
.
4.
Connect the
7
handheld controller to the
9
power controller using the
8
a
auxiliary cable.
NOTE:
The
7
handheld controller may also be connected directly to the
5
umbilical using the
8
b
auxiliary cable.
5.
Mount the appropriate 3-axis nozzle configuration to the swivel mount of the
crawler.
6.
Connect the
4
3-axis encoder cable to the encoder cables of the
2
encoded skew vertical probe holder and the
1
slider pps encoder.
7.
Connect the opposite end of the
4
3-axis encoder cable to the
6
user’s
instrument and to the
9
power controller.
CAUTION!
DO NOT DISCONNECT
UNDER LOAD.
Shut off power before connection or
disconnecting. Permanent damage to electronics
could occur.