JK2830 Operation Manual Chapter 5 Execute LCR operation and some examples
correction, the short plate of low impedance should be connected between test
terminals. Another way is to directly connect Hc with Lc or Hp with Lp, then
connect both.
Note:
When the DUT is a polarity component, before testing, the high potential
terminal should be connected to the terminal with mark “+”, “Hc” or “Hp” and the
low terminal should be connected to the terminal with mark
“
-
”, “Lc” or “Lp”.
Warning: Before testing, please discharge the tested polarity component
so as to avoid the damage to the instrument.
5.2
Eliminate the influence of stray impedance
Figure 5-2 Eliminate the influence of stray capacitance
When the DUT has high impedance(such as small capacitance), the influence of
stray capacitance cannot be ignored. Figure 5-1 is an example of the use of 4
terminal pair measurement. In this figure, Cd is connected with Cx in a parallel
way and when a conductance plate is positioned under DUT, capacitance Ch will
connect with Cx in parallel after connecting with Cl in series and by this way the
Hp
Lp
Lc
Hc
Figure 5-1 Influence of stray capacitance
Test terminal
Shielding
ground
Metal
conductor
Cx
Cd
Ch
Cl
Cx
Hc
Hp
Lp
Lc
Shielding
plate
Shielding
ground
Test terminal
Metal
conductor
50