PDL5 User Manual
PDL5-UM-00001 Rev 001
60
*WAI Command
Syntax
*WAI
Function
Prevents the meter from executing any further commands or
queries until all previously pending operations have been
completed.
There are no consequences to this command because all commands
are executed sequentially; therefore, any subsequent commands
are completed by the time this command is parsed.
Example
*WAI
SCPI Commands
[:SOURce]:WAVlength
Syntax
[:SOURce]:WAVlength<space>[wavelength|MIN|MAX|DEF] OR
[:SOURce]:WAVlength[:NEXT]
Function
Switches to a specified or next available wavelength.
no parameter or [:NEXT]:
switches to the next available
wavelength
wavelength:
switches to the specified wavelength in nm. Other
units can be specified
MIN, MAX or DEF:
switches to the first, last or default
wavelength respectively.
Example
WAV 1.55 um
switches the meter’s output to 1550 nm.