BA-40100-02-V17 | 33
Operating instructions
mg/l
mV
pH
∞C
ESC
039 702 719 63
2
ab<=> 01
Set_2
preselects
1.4.8.1
edit name and save
ab<=> AB
save
Fig. 33: Main menu >> status >> presettings >> apply
Enter the name of the pre-adjustment using the arrow keys
▲
and
▼
.
This means:
Keys
Function
▲
and
▼
:
Change individual letters and numbers
◀
and
▶
:
shift cursor
ab <=> 01 change small letters to numbers/symbols
ab <=> AB change small into capital letters
Press the "save" key to save up to this point the adjusted regulating
characteristics. Confirm the procedure with the “OK” key.
The following adjustments are saved and can be re-activated at any
time:
• regulating characteristics of configured controller
• setpoints of configured controller
19.1.7 Menu 2.1: adjust controller
mg/l
mV
pH
pH value
disinfection
ESC
039 702 719 09
0
setup
comb. cl.
% salt
controller
2.1
mg/l
Fig. 34: Main menu >> Menu 2 >> controller
The TOPAX DX is able to command up to four control units (see chap-
ter “8 Controller explanation” on page 24). The basic adjustments of
every single control unit are configured in this menu:
Menu
Controller setting
Required specifications
2.1.1
Disinfection
2.1.1.1
Direction
boosting/dropping/2 sites
2.1.1.2
Disturbance variable
Disturbance variable, ON/OFF
2.1.1.3
Overview of settings
2.1.1.3.1
Parameter Xp
%-value
2.1.1.3.2
Parameter Tn
ON or OFF + time
2.1.1.3.3
Parameter Tv
ON or OFF + time
2.1.1.4
Basic load
ON or OFF + time
2.1.1.5
Manual mode
%-value, ON/OFF, shock
chlorination
2.1.2
pH value
2.1.2.1
Direction
boosting/dropping/2 sites
2.1.2.2
Disturbance variable
Disturbance variable, ON/OFF
2.1.2.3
Overview of settings
2.1.2.3.1
Parameter Xp
%-value
2.1.2.3.2
Parameter Tn
ON or OFF + time
2.1.2.3.3
Parameter Tv
ON or OFF + time
2.1.2.4
Basic load
ON or OFF + time
2.1.2.5
Manual mode
%-value, ON/OFF
2.1.3
combinded chlorine
2.1.3.1
Direction
boosting/dropping/2 sites
2.1.3.2
Disturbance variable
Disturbance variable, ON/OFF
2.1.3.3
Overview of settings
2.1.3.3.1
Parameter Xp
%-value
2.1.3.3.2
Parameter Tn
ON or OFF + time
2.1.3.3.3
Parameter Tv
ON or OFF + time
2.1.3.4
Basic load
ON or OFF + time
2.1.3.5
Manual mode
%-value, ON/OFF
2.1.4
controller for salt content
2.1.4.1
Direction
boosting/dropping/2 sites
2.1.4.2
Disturbance variable
Disturbance variable, ON/OFF
2.1.4.3
Parameter
2.1.4.3.1
Parameter Xp
%-value
2.1.4.3.2
Parameter Tn
ON or OFF + time
2.1.4.3.3
Parameter Tv
ON or OFF + time
2.1.4.4
Basic load
ON or OFF + time
2.1.4.5
Manual mode
%-value, ON/OFF
Select one of the four control units. The device goes to the
relevant Menu 2.1.1-2.1.4.
Apply the adjustments relevant to your requirements in the
following sub-menus to the control unit.
Adjust the parameters accordingly with the arrow keys
▲
and
▼
and confirm your entries with the “OK” key.