Parameter setup and measuring operation
20
LDM52 OEM, Release 003, Revision 000
6.4.2
BR – Baud rate
BR enables the adjustment of the serial baud rate x.
As soon as a new baud rate is set, the device will start communicating with the new baud rate.
BR will not be modified upon a parameter reset via PR.
Prior to setting a high baud rate of > 115200 baud, make sure that the subsequent sys-
tem is capable of processing that baud rate.
Any check of the set baud rate is impossible after wrong input of the baud rate (differ-
ence between LDM52 OEM and communication program).
Output: Baud rate of serial port [BR]: 115200
6.4.3
HE – Heating adjustment
The parameter HE defines the switching thresholds for switching an external heating element on
and off.
The command is enabled only where the device is actually equipped with a heating element.
Query:
HE
Set:
HEx y
Range of parameter x:
Switching on heating
-40 … 40 (integer)
Range of parameter y:
Switching off heating
-40 … 40 (integer)
Standard:
HE4 10
For switching the heating on or off, the internal measured temperature is compared to the set
parameters.
Internal temperature < x (HeatON)
Heating is switched on.
Internal temperature > y (HeatOFF)
Heating is switched off.
Please observe the following when setting the parametrization: x (HeatON) <= y (HeatOFF).
Query:
BR
Set:
BRx
Range of parameter x:
1200,2400,4800,9600,14400,19200,28800,38400,56000,
57600,115200,128000,230400,256000
Standard:
115200 baud/ 8 data bits /1 stop bit / no parity