Product Information / Technical Service Manual 

JEDMED PHOENIX III Treatment Chair

 

 

JEDMED INSTRUMENT COMPANY 

5416 JEDMED CT. 

ST. LOUIS, MO  63129-2217 

PHONE (314) 845-3770 

FAX (314) 845-3771 

Email [email protected] 

 

DOC. 9810 

REV. F, 12/05/14 

L'information de produit/manuel d'entretien technique technique 

Chaises de traitement de la série JEDMED Phoenix III 

 

IMPORTANT:  Read all instructions before assembling or operating this device. 

An E

mplo

yee-O

wned

 Com

pany

 

 

Содержание PHOENIX III

Страница 1: ... ST LOUIS MO 63129 2217 PHONE 314 845 3770 FAX 314 845 3771 Email info jedmed com DOC 9810 REV F 12 05 14 L information de produit manuel d entretien technique technique Chaises de traitement de la série JEDMED Phoenix III IMPORTANT Read all instructions before assembling or operating this device An Employee Owned Company ...

Страница 2: ...IPMENT MIRROR LIGHT SHADE REMOVAL LIGHT RE LAMPING FOOT CONTROL INSTALLATION OPERATION OPTIONAL EQUIPMENT CLEANING INFECTION CONTROL MAINTENANCE SERVICE FACILITIES ELECTRICAL COMPONENTS DIAGRAM 115VAC 230VAC ELECTRICAL SCHEMATIC 115VAC 230VAC ELECTRICAL COMPONENTS PART NUMBER LISTING EXPLODED VIEW FRAME PARTS LISTING FRAME EXPLODED VIEW BASE PARTS LISTING BASE EXPLODED VIEW ARM UPHOLSTERY INSTALLA...

Страница 3: ... ROTATION MUST BE LOCKED BEFORE ALLOWING THE PATIENT TO ENTER OR EXIT THE CHAIR A NEARLY FLAT RECLINE POSITION IS PROVIDED FOR MINOR SURGICAL APPLICATIONS SPECIALLY DESIGNED WINGS ON THE BACKREST PROVIDES SUPPORT FOR THE PATIENT S ARMS COMFORTABLY AT THEIR SIDES AN ERGONOMICALLY DESIGNED LUMBAR SUPPORT ADDS TO THE PATIENT S COMFORT WHILE KEEPING THE HEAD PROPERLY POSITIONED FOR EASY ACCESS THE ARM...

Страница 4: ...CAN BE RAISED TO A MAXIMUM ELEVATION AND THEN ROTATED 355 DEGREES WITHOUT TIPPING THE CHAIR WARNING UNSUPERVISED CHILDREN SHOULD NEVER BE ALLOWED IN THE VICINITY OF THE CHAIR OR ANY ELECTRICAL DEVICES IT IS THE RESPONSIBILITY OF THE PHYSICIAN AND THE STAFF TO ENSURE THE PROPER AND SAFE USE OF THIS CHAIR JEDMED Instrument Company was formed in July 1978 and is entirely employee owned ESOP Our goal ...

Страница 5: ...IT POSITION 3 LOW VOLTAGE D C MOTORS ARE STANDARD D C MOTORS ARE SAFE QUIET AND HAVE A LONGER LIFE EXPECTANCY APPLICATION OF PRODUCT THIS PRODUCT IS DESIGNED FOR THE PURPOSE OF COMFORTABLY POSITIONING PATIENTS DURING MEDICAL EXAMINATIONS AND PROCEDURES CHAIR SECTIONS MOVE TO ALLOW THE MOST ADVANTAGEOUS WORKING POSITION FOR THE DOCTOR AND HIS HER ASSISTANT S REV F 12 05 14 PRODUCT REQUIRES INITIALI...

Страница 6: ...TURE DU MANUEL DE L OPÉRATEUR MAIN POWER ON MAINS POWER ON MAIN POWER OFF MAINS POWER OFF PROTECTIVE EARTH GROUND PROTECTION ÉCHOUEMENT TERRE TYPE B EQUIPMENT ÉQUIPEMENT DE TYPE B EARTH GROUND ÉCHOUEMENT TERRE NORMAL OPERATION TEMPERATURE RANGE STORAGE SHIPPING TEMPERATURE RANGE INDOOR USE ONLY USAGE INTÉRIEUR SEULEMENT FONCTIONNEMENT NORMAL Plage de température STORANGE ET EXPÉDITION GAMME DE TEM...

Страница 7: ...matériel de transmissions portatif et mobile de rf peut effectuer l appareillage électrique médical 3 L utilisation des accessoires des capteurs et des câbles autres que ceux indiqués par le fabri cant peut avoir comme conséquence les plus grandes émissions ou l immunité diminuée de la chaise de PHOENIX III 4 La chaise PHOENIX III ne devrait pas être employée à côté de ou empilée avec l autre équi...

Страница 8: ... RÉFÉRENCE 4 HEED WARNINGS ADHERE TO ALL WARNINGS ON EQUIPMENT AND IN OWNER S MANUAL AVERTISSEMENTS D ATTENTION ADHÉREZ À TOUS LES AVERTISSEMENTS SUR L ÉQUIPEMENT ET DANS LE PROPRIÉTAIRE S MANUEL 5 ACCESSORIES DO NOT USE ANY ACCESSORIES NOT RECOMMENDED BY THE MANUFACTURER AS THEY MIGHT CAUSE HAZARDS ACCESSOIRES N UTILISEZ AUCUN ACCESSOIRE NON RECOMMANDÉ PAR FABRICANT COMME ILS POURRAIENT CAUSER DE...

Страница 9: ...LED INTO EQUIPMENT OBJECTS HAVE FALLEN INTO EQUIPMENT EQUIPMENT HAS BEEN EXPOSED TO RAIN OR WATER EQUIPMENT HAS BEEN DROPPED OR DAMAGED EQUIPMENT SHOWS SIGNIFICANT CHANGE IN OPERATION EQUIPMENT DOES NOT OPERATE NORMALLY DOMMAGES OU ÉCHECS EXIGEANT LE SERVICE DÉBRANCHEZ L ÉQUIPEMENT DE LA SOURCE D ÉNERGIE ET RÉFÉREZ VOUS AUX PERSONNES QUALIFIÉES DE SERVICE SI L UN DES SUIVRE SE PRODUISENT LA CORDE ...

Страница 10: ... ETC BASE AUTO STOP BASE DOWN FUNCTION WILL AUTOMATICALLY STOP IF AN OBSTRUC TION IS DETECTED UNDER THE BASE LIFTING ARM OBSTRUCTION MUST BE REMOVED BEFORE ANY CHAIR FUNCTIONS CAN BE REACTIVATED 15 ALWAYS INSURE THAT THE UPHOLSTERY IS CORRECTLY ATTACHED TO THE CHAIR FRAME BEFORE USING THE CHAIR 16 ALWAYS DISCONNECT POWER BEFORE WORKING ON SERVICING THE CHAIR TOUJOURS S ASSURER QUE LA SELLERIE EST ...

Страница 11: ...N MODE INTERMITTENT DUTY INDOOR USE ONLY 3MIN 17 MIN INDOOR USE ONLY OPERATING 5ºC TO 40ºC 41ºF TO 104º F RELATIVE HUMIDITY 20 to 90 30ºC TRANSPORTATION NOT CONDENSING TEMPERATURE STORAGE 10ºC TO 50ºC 14ºF TO 122º F RELATIVE HUMIDITY 20 to 90 30ºC TEMPERATURE NOT CONDENSING WEIGHT CAPACITY 500 lbs 228kg SUBJECT TO NORMAL WEIGHT DISTRIBUTION NOTE DO NOT SIT ON BACK ARMS OR TOE SECTION BACK ARM AND ...

Страница 12: ...X 6 HEADREST ADJUSTMENT PRODUCT SPECIFICATIONS PHOENIX III REV F 12 05 14 AGENCY Standard for Medical Electrical Equipment Part 1 General Requirements for Safety UL 60601 1 1st Edition Dated 04 25 2003 with revisions through 04 26 2006 and CAN CSA C22 2 No 601 1 M90 11 03 ETL 46979 ELECTRICAL 115VAC 60HZ 5A 230VAC 50 60HZ 5A ...

Страница 13: ...AGRAM TO THE RIGHT IS SHOWING THE CHAIR IN THE BACK FULL DOWN AND BASE FULL DOWN POSITION THE CHAIR TOP SECTION WILL MOVE EIGHT INCHES TOWARDS THE HEAD OF THE CHAIR AS THE BASE IS RAISED DUE TO THIS MOVEMENT MORE CLEARANCE IS REQUIRED AT THE HEAD OF THE CHAIR AN 8 FOOT DIAMETER CIRCLE AREA IS REQUIRED TO INSURE THAT THE CHAIR WILL NOT HIT OBJECTS WHEN IT IS ROTATED IN EITHER THE BASE DOWN OR BASE ...

Страница 14: ...CHAIR AND BASE ASSEMBLED WITH UPHOLSTERY HEADREST FOUR ANCHORS FOR ANCHORING TO FLOOR ACCESSORIES OWNER S OPERATOR S MANUAL NOTE ACCESSORIES MAY BE PACKAGED SEPARATELY POWER REQUIREMENTS CHECK CHAIR IDENTIFICATION LABEL LOCATED ON ELECTRICAL PANEL FOR REQUIRED VOLTAGE SPECIFICATIONS CONNECT CHAIR TO HOSPITAL GRADE RECEPTACLE ONLY Vérifier l étiquette d IDENTIFICATION PRÉSIDENCE situés sur le panne...

Страница 15: ...ROUGH MOUNTING HOLES ANCHOR BASE TO FLOOR WITH FOUR ANCHORS SCREWS PROVIDED 2 MAKE UTILITY CONNECTIONS AS REQUIRED 3 INSURE ALL SAFETY COVERS ARE IN PLACE AND REMOVABLE UPHOLSTERY IS SECURE SHIFTING MAY HAVE OCCURRED DURING HANDLING OR SHIPPING 4 PLUG CHAIR INTO REQUIRED VOLTAGE OUTLET REFER TO CHAIR IDENTIFICATION LABEL LOCATED ON THE ELECTRICAL PANEL FOR PROPER CHAIR VOLTAGE REQUIREMENT NOTE CHA...

Страница 16: ...PROGRAMMED CORRECTLY 5 REPEAT STEPS 2 3 AND 4 UNTIL ALL DESIRED POSITIONS ARE PROGRAMMED YOU CAN SAVE UP TO 3 POSITIONS 6 AUTO RETURN IS A PRESET POSITION AND CANNOT BE CHANGED THE PROGRAMMED POSITIONS CAN BE ATTAINED AT ANYTIME AND FROM ANY POSITION GENERAL OPERATING INSTRUCTIONS 1 SEE FOLLOWING PAGES OF THIS SECTION FOR SYMBOLS AND THEIR DEFINITIONS 2 WHILE SAFETY FEATURES PREVIOUSLY COVERED HAV...

Страница 17: ...esired position Press and HOLD the SAVE button until a BEEP is heard about 5 seconds Release the SAVE button and press and release button 1 2 or 3 Repeat procedure to override a previous position RESET PROCEDURE Press and hold position 1 and position 2 buttons at the same time Chair should beep about 10 times and then stop beeping After beeping stops release buttons 1 and 2 and press and release t...

Страница 18: ...D POSITION CAUTION IT IS RECOMMENDED THAT THE HEADREST HEIGHT IS ADJUSTED BEFORE LOWERING THE CHAIR BACK IL EST RECOMMANDÉ QUE LA TÊTE DE LA HAUTEUR est ajustée avant d abaisser le dossier du fauteuil WHEN HEADREST ADJUSTMENT IS NECESSARY WITH PATIENT SEATED BE SURE TO SUPPORT THE PATIENT S HEAD AND NECK AT ALL TIMES WHILE ADJUSTMENT IS BEING DONE QUAND TÊTE AVEC ajustement est nécessaire patient ...

Страница 19: ... POSITION LOCKED TENSION ADJUSTMENT PULL LEVER OUT TO THE UNLOCKED POSITION TURN TENSION BAR SCREW 1 8 TURN CLOCKWISE TO INCREASE TENSION USING A 3 16 ALLEN WRENCH PUSH THE LEVER IN AND CHECK FOR DESIRED HOLDING TENSION REPEAT AS NECESSARY 1 2 3 4 NOTE la tête du patient de soutien tout en faisant des ajustements FLIP LOCKING LEVER AWAY FROM THE BODY OF THE MECHANISM TO ALLOW HEADREST ARTICULATION...

Страница 20: ...ON OR IT IS AT FULL DOWN POSITION TO RAISE THE ARM RAISE ARM FULLY THE ARM AUTOMATICALLY LOCKS IN THE UPRIGHT POSITION WHEN ROTATED TO THE FULL UPRIGHT POSITION IT IS NOT NECESSARY TO DEPRESS THE ARM RELEASE TAB TO RAISE THE ARM CAUTION DO NOT SIT ON ARMS OR USE THEM TO LIFT OR MOVE CHAIR DOING SO MAY CAUSE DAMAGE TO PRODUCT AND OR BODILY INJURY Ne pas s asseoir sur les armes ou les utiliser pour ...

Страница 21: ...WER CORD TO THE CHAIR DISCONNECT THE CABLE TO THE RIGHT SIDE MEMBRANE KEY PAD AND PULL THE CABLE BACK THROUGH THE CENTER BACK FRAME THIS IS TO PREVENT DAMAGING THE CABLE WHEN DRILLING THROUGH THE MOUNTING HOLES REF PIC A REPLACEMENT BULB JEDMED P N 99 4160 NEW REPLACEMENT BULB SHOULD NOT BE TOUCHED USE A PAPER TOWEL OR CLOTH TO HOLD REPLACEMENT BULB A B C REMOVE THE CHAIR BACK UPHOLSTERY OPERATION...

Страница 22: ...ill expose the bulb 5 Gently pull the bulb forward releasing it from the socket 6 Use plastic wrapper or tissue paper as protection for the new bulb when replacing Oils or impurities from fingers can adversely affect the life of Quartz halogen bulbs The lamp is furnished with a 12 Volt 20 Watt Halogen bulb G4 Pin Base OSRAM 64425 or Phillips 12452G4 or equivalent 7 Insert the pins of the new bulb ...

Страница 23: ...L CONNECT FOOT CONTROL CABLE FOOT CONTROL FUNCTIONS OPERATE THE SAME AS THE MEMBRANE SWITCHES ON THE BACK OF THE CHAIR SEE PAGE 17 BASE MUST BE PRE WIRED AT THE FACTORY FOR THE FOOT CONTROL OPTION REMARQUE Base de mut être précâblés en usine pour l option commande au pied REV F 12 05 14 UP UP DOWN DOWN BASE BACK ...

Страница 24: ...E IN CONTACT WITH PATIENT SHOULD BE INSPECTED PRIOR TO EACH USE GFCI OUTLETS AND BREAKERS SHOULD BE TESTED ACCORDING TO MANUFACTURERS RECOMMENDATIONS LE SERVICE INFORMATIQUE EST RECOMMANDÉ QUE LA CHAISE ET L ÉQUIPEMENT SOIENT COMPLÈTEMENT NETTOYÉS ET DÉSINFECTÉS APRÈS CHAQUE PATIENT EMPLOYEZ SEULEMENT LES DÉCAPANTS ET LES DÉSINFECTANTS APPROUVÉS POUR LE VINYLE DE NETTOYAGE N EMPLOYEZ PAS LES DÉCAP...

Страница 25: ...PAD HAND CONTROL SWITCH HARNESS LINAK JUNCTION BOX BASEACTUATOR LINAK CONTROL BOX POWER CABLE A C INLET HARNESS A C POWER CABLE A C EXTENSION CABLE BACK LIGHT HARNESS LIGHTOPTION ELECTRICAL DIST BOX SAFETY SWITCH HARNESS EXTENSION CABLE LIGHT FUSE CIRCUIT BREAKER SWITCH 21 18 4 20 1 19 15 10 13 8 2 11 12 SAFETY SWITCH 14 3 9 17 24 16 LOCK CLIP 6 ...

Страница 26: ...LE FOOT HAND CONTROL CABLE SWITCH PAD HAND CONTROL SWITCH HARNESS LINAK JUNCTION BOX BASEACTUATOR LINAK CONTROL BOX POWER CABLE A C INLET HARNESS A C POWER CABLE A C EXTENSION CABLE BACK LIGHT HARNESS LIGHTOPTION ELECTRICAL DIST BOX SAFETY SWITCH HARNESS EXTENSION CABLE LIGHT FUSE CIRCUIT BREAKER SWITCH 21 18 4 20 1 19 15 10 13 8 2 11 12 SAFETY SWITCH 14 3 9 17 24 5 LOCK CLIP 16 7 6 ...

Страница 27: ... NO COM NC NO COM 21 1 1 1 1 1 1 2 HB 4 3 2 1 M 1 M 1 SW9 SW3 SW2 SW6 SW5 SW8 SW7 SW1 SW4 SW10 1 1 1 123 SW18 LD1 LD2 LN2 LN1 123 1234 1 2 L 18 6 4 1 19 20 15 12 7 10 13 8 9 2 TR1 17 11 24 5 OPTIONAL TEECABLE FOOTCONTROL OPTIONAL COILED CABLE FOOTCONTROL OPTIONAL FOOTCONTROL 3 ...

Страница 28: ... COM NC NO COM 21 1 1 1 1 1 1 2 HB 4 3 2 1 M 1 M 1 SW9 SW3 SW2 SW6 SW5 SW8 SW7 SW1 SW4 SW10 1 1 1 123 SW18 LD1 LD2 LN2 LN1 123 1234 1 2 L 18 6 4 1 19 20 15 12 7 10 13 8 9 2 TR1 17 11 24 5 OPTIONAL TEECABLE FOOTCONTROL OPTIONAL COILED CABLE FOOTCONTROL OPTIONAL FOOTCONTROL 3 R1 R1 ...

Страница 29: ...30V LINAK DJB MODULE FOOT CONTROL 2 PEDAL T CABLE LINAK CHAIR WITH FOOT CONTROL OPTION STRAIGHT CABLE CHAIRS WITHOUT FOOT CONTROL COILED CABLE FOOT CONTROL AC POWER CORD 115VOLT AC POWER CORD 230V EURO AC POWER CORD 230V ARGENTINA AC POWER CORD 230V UK IRELAND AC POWER CORD 230V ITALY FUSE 600mA 115V CHAIRS FUSE 300mA 230V CHAIRS CABLE CB6 POWER CABLE LIGHT CABLE SAFETY SWITCH CABLE AC POWER INLET...

Страница 30: ...28 9 48 49 53 48 54 50 LIGHTLY GREASE 62 1720 ON MOVING MATING SURFACES LIGHTLY GREASE 62 1720 ON MOVING MATING SURFACES 26 60 48 47 55 8 50 WIRETROUGH AND SUPPORT IN POSITION 17 16 15 18 35 39 43 23 REARISOMETRIC VIEW 30 29 DETAILA SCALE1 4 45 19 26 33 20 21 34 17 8 26 24 25 51 TOP VIEW DETAILB SCALE1 4 40 59 26 13 ...

Страница 31: ...31 PARTS BREAKDOWN PHOENIX III CHAIR FRAME ASSEMBLY REV F 12 05 14 41 23 38 37 60 48 43 23 37 60 48 60 48 6 ELECTRICALCOMPONENTLAYOUT 57 46 4 3 36 2 61 60 48 35 39 24 35 39 2 61 22 48 14 56 ...

Страница 32: ...5 SELF ADHESIVE 29 1 55 0311 VELCRO HOOK 1 X 16 SELF ADHESIVE 30 2 55 0302 VELCRO HOOK 5 8 X 6 SELF ADHESIVE 31 4 59 1861 HEX 3 8 16 UNC 2A X 1 1 4 32 4 59 1863 HEX 3 8 16 UNC 2A X 1 3 4 33 3 59 1988 SHOULDER 3 8 x 5 8 STAINLESSSTEEL 34 2 59 1981M SHOULDER 3 8 x 5 8 MODIFIED THREAD 35 2 59 1984 SHOULDER 10mm X 45mm 36 4 59 2006 NUT 6 32 UNC W STAR WASHER 37 2 59 2013 HEX NUT 1 4 20 UNC 38 1 59 201...

Страница 33: ...3 84 86 85 11 46 71 57 39 66 16 38 61 62 14 29 22 34 73 33 45 72 66 64 25 24 23 28 48 82 18 45 72 40 66 21 10 79 76 13 62 36 49 68 40 17 67 51 62 39 62 89 87 8 15 50 75 83 32 35 65 91 37 90 83 32 30 31 62 70 74 81 88 DETAILA SCALE1 4 47 78 47 55 43 53 43 53 77 77 47 47 43 44 42 53 43 44 44 41 42 53 20 19 26 27 69 ...

Страница 34: ...WASHER FLAT NYL 5 8 X 1 X 1 4 THK 45 9 59 1861 HEX 3 8 16 UNC 2A X 1 1 4 46 4 59 1875 HEX 1 4 20 UNC 2A X 3 4 ITEM NO QTY PARTNO DESCRIPTION 47 8 59 1880 HEX 5 8 18 UNF 2A X 2 1 2 GR8 48 2 59 1985 SHOULDER 1 4 X 3 8 SST 49 2 59 2006 NUT 6 32 UNC W STAR WASHER 50 4 59 2012 HEX NUT 10 24 UNC 51 1 59 2013 HEX NUT 1 4 20 UNC 52 1 59 2025 HEX NUT 3 8 16 UNC 53 8 59 2030 HEX NUT 5 8 18 UNF NYLOC THIN 54...

Страница 35: ...ACK DELRIN ARM OPTION 9 1 91 2801R WELDMENT ARMREST RH 8 1 91 2801L WELDMENT ARMREST LH 7 2 60 1571 SPRING COMP 343 X 1 25 X 035 6 2 91 2389 ARM LOCK RELEASEBAR M3000 5 2 90 2388 ARM LOCK REAR RELEASE 4 1 93 2387R HOUSING RH ARM LOCK M3000 3 1 93 2387L HOUSING LH ARM LOCK M3000 2 1 91 2386R SEATHINGE MACHINED W ARM OPTION RH 1 1 91 2386L SEATHINGE MACHINED W ARM OPTION LH ITEM NO QTY PARTNO DESCRI...

Страница 36: ...NTO THE SEAT FRAME WITH THE FRONT UP SLIGHTLY SLIDE THE UPHOLSTERY BACK UNTIL THE TWO REAR BRACKETS ENGAGE THE BACK OF THE SEAT FRAME THERE ARE TWO SLOTS IN THE SEAT FRAME WHERE THESE BRACKETS MUST ENGAGE THEN DROP THE FRONT OF THE UPHOLSTERY DOWN UNTIL THE RETAINER CLIP ENGAGES THE SLOT IN THE FRONT OF THE SEAT FRAME REAR BRACKETS SLOTS IN BACK OF SEAT FRAME FRONT RETAINER CLIP LEG UPHOLSTERY P N...

Страница 37: ...HOLSTERY HEADREST ENT ITEM 1 67 1340U ISINTERCHANGEABLEWITH 67 1361U 1 ARM PADSNOTINCLUDEIN PACKAGE ARM PAD P N K905013 2 ITEM NO QTY PARTNO DESCRIPTION 4 1 67 4451 UPHOLSTERY LEG SUPPORT 3 1 67 3011 UPHOLSTERY SEAT 2 1 67 4450 UPHOLSTERY BACK 1 1 67 1360U UPHOLSTERY HEADREST ...

Страница 38: ... 8 TURN CLOCKWISE TO INCREASE BRAKE HOLDING ACTION CHECK THE BRAKE HOLDING ACTION AND REPEAT IF NECESSARY REV F 12 05 14 UTILISEZ UN TOURNEVIS PHILLIPS DU NUMÉRO 2 POUR SERRER LES DEUX VIS D APPROCHE DES VIS D APPROCHE PEUVENT ÊTRE ACCÉDÉES SANS ENLEVER LA COUVERTURE BASSE FAITES LES AJUSTEMENTS AVEC LE FREIN EN MODE LIBÉRÉ SERREZ CHAQUE TOUR DE LA VIS 1 8 DANS LE SENS DES AIGUILLES D UNE MONTRE P...

Страница 39: ...LY CAUSED BY SOMEONE SITTING ON THE ARM OR LIFTING THE CHAIR BY THE ARM HEADREST WILL NOT JAMMED LOCK MECHANISM PULL HEADREST OUTWARD WITH RELEASE BOTH HANDS IF THIS PROBLEM PERSISTS CALL FOR SERVICE MEMBRANE SWITCHES CHECK THAT CABLE CONNECTOR CN30 TO THE SWITCH PAD IS TIGHT DON T WORK DEFECTIVE MEMBRANE SWITCH PAD REPLACE BASE SWIVEL LOCK ADJUST TENSION SCREWS 1 8 TURN EACH AND CHECK LOCKING WON...

Страница 40: ... by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol P d 2 1 P d 2 1 P d 3 2 NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflectio...

Страница 41: ...ommon mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 d...

Страница 42: ...wer rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from struc tures objects and people Guidance and manufacturer s declaration electromagnetic emission...

Страница 43: ...43 JEDMED Instrument Company 5416 Jedmed CT St Louis MO 63129 2217 Phone 314 845 3770 Fax 314 845 3771 REV F 12 05 14 ...

Отзывы: