Chapter 10
Measure Mode – Spectrum Tab
Upstream Spectrum
250
DSAM Help User Guide
Rev. 001
5
Press the
Limits
softkey.
6
Using the arrow keys, select
Off
or a limit set (
Tap
,
Ground
Block
,
TV
,
Cable Modem
,
Proof
, or
custom limits 1-3
).
7
Press
ENTER
.
8
To edit the limit value, press the
Limits
softkey.
9
Using the arrow keys, select the
Edit
option.
10
Press
ENTER
.
Understanding
Upstream
Spectrum Results
The graph continuously updates as each frequency is measured. This
forms a graph across the entire frequency range. When excessive
ingress noise occurs at a specific frequency, the measurement line
develops a peak on the graph.
NOTE
When you select Edit, the limit-set parameters you define remain
constant for all channel plans and measurement modes.
Figure 227
Configure – Select a Limit Set to Edit
NOTE
See
for instructions on editing limit sets.
If you chose the wrong limit set before conducting the test, press
the
Limits
softkey and select a different limit set to view your test
results in relation to a different limit set without rerunning the test.
21128026 R001 DSAM Help User Guide.book Page 250 Thursday, February 5, 2009 2:02 PM
Содержание DSAM-1500
Страница 2: ...21128026 R001 DSAM Help User Guide book Page ii Thursday February 5 2009 2 02 PM ...
Страница 4: ...21128026 R001 DSAM Help User Guide book Page iv Thursday February 5 2009 2 02 PM ...
Страница 361: ...21128026 R001 DSAM Help User Guide book Page 321 Thursday February 5 2009 2 02 PM ...