JD780A Series Analyzers
470
Using Mobile WiMAX Signal Analyzer
Conducting power vs. time (frame) measurements
The Power vs. Time (Frame) measures the modulation envelope in the time domain, showing the
power of each time slot in a Mobile WiMAX signal.
Setting measure setup
After configuring test parameters as described in the
“Configuring test parameters” on page 456, you
can continue your measurement. The measurement settings can be saved and recalled as a file.
Procedure
1.
Press the
MEASURE SETUP
hot key.
2.
Press the
System Config
soft key, and then do the following:
a.
Press the
Bandwidth
soft key, and then select the nominal channel bandwidth to be
measured from the following choices:
–
7 MHz (8/7):
Sets the channel bandwidth to 7 MHz with sampling factor 8/7. In
conjunction with nominal channel bandwidth
–
8.75 MHz (8/7):
Sets the channel bandwidth to 8.75 MHz with sampling factor 8/7
–
10 MHz (28/25):
Sets the channel bandwidth to 10 MHz with sampling factor 28/25
b.
Toggle the
Frame Length
soft key and select
5 ms
or
10 ms
.
c.
Press the
TTG (RTG)
soft key to set the transmit/receive transition gap between the last
sample of the downlink burst and the first sample of the subsequent uplink burst.
d.
Enter a value by using the numeric keys, and then press the
Enter
soft key.
e.
Press the
CP Ratio
soft key, and then select the cyclic prefix ratio option:
1/4
,
1/8
,
1/16
, or
1/32
.
f.
Press the
DL Symbols
soft key to set the number of downlink symbols.
g.
Enter a value by using the numeric keys, and then press the
Enter
soft key.
h.
Press the
UL Symbols
soft key to set the number of uplink symbols.
i.
Enter a value by using the numeric keys, and then press the
Enter
soft key.
3.
To set the start/stop symbols, complete the following steps:
a.
Press the
Start Symbol
soft key.
b.
Enter a value by using the numeric keys, and then press the
Enter
soft key.
c.
Press the
Stop Symbol
soft key.
d.
Enter a value by using the numeric keys, and then press the
Enter
soft key.
4.
Toggle the
Preamble Index
soft key and select the preamble search method:
Auto
or
Manual
.
5.
Toggle the
Search Type
soft key and select
Full
or
Window
.
6.
Toggle the
Video Filter
soft key and select
On
or
Off
to enable or disable the video filter.
7.
Optional
. Press the
Miscellaneous
soft key, and then do the following as needed:
–
To set the delay, complete the following steps:
a.
Press the
Delay
soft key to set the amount of delay in µs.
b.
Enter a value by using the numeric keys. You can also use the rotary knob.
c.
Press the
µs
soft key.
NOTE
The Delay setting is used only when there is a time offset in the signals to be measured.
Содержание CellAdvisor JD780A Series
Страница 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Страница 91: ...JD780A Series Analyzers Using Spectrum Analyzer 91 Figure 23 PIM detection for multiple carriers...
Страница 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Страница 525: ......
Страница 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...