JD780A Series Analyzers
354
Using CDMA Signal Analyzer
Procedure
1.
Press the
TRACE/DISPLAY
hard key.
2.
Press the
Code Order
soft key, and then select the code order option from the following choices:
–
Hadamard:
To assign the number based on Walsh code order and display the measured
code power according to the assigned number in sequence.
–
Bit Reverse:
Code number in binary bit is reversed relative to the Hadamard method.
3.
Toggle the
Base Code Length
soft key and select the base Walsh code number:
64
or
128
.
CDP Table
The CDP Table is used to look at a numeric and textual description of all the active code channels in
the signal.
Setting measure setup
After configuring test parameters as described in the
“Configuring test parameters” on page 334, you
can continue your measurement. The measurement settings can be saved and recalled as a file. You
can also use JDViewer, PC application software to configure a measure setup, save as a file, and load
the file on to the instrument.
Procedure
1.
Press the
MEASURE SETUP
hot key.
2.
To set the Pseudo-Noise (PN) offset, complete the following steps:
a.
Toggle the
PN Offset
soft key between
Auto
and
Manual
.
b.
Enter a value by using the numeric keys to set manually.
c.
Press the
Enter
soft key.
3.
To set the incremental step for the PN offset, complete the following steps:
a.
Press the
PN Increment
soft key.
b.
Enter a value by using the numeric keys.
c.
Press the
Enter
soft key.
4.
To set the threshold between Auto and Manual, complete the following steps:
a.
Press the
Threshold
soft key. Its default setting is -27 dB.
b.
Enter a value between -50 and zero by using the numeric keys to change the value in the
Manual mode.
c.
Press the
Enter
soft key.
5.
Toggle the
Q-Paging
soft key between
On
and
Off
to enable or disable the quick page channel
that indicates the power of Quick Page Walsh code 80.
Содержание CellAdvisor JD780A Series
Страница 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Страница 91: ...JD780A Series Analyzers Using Spectrum Analyzer 91 Figure 23 PIM detection for multiple carriers...
Страница 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Страница 525: ......
Страница 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...