![JDS Uniphase CellAdvisor JD780A Series Скачать руководство пользователя страница 280](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616280.webp)
JD780A Series Analyzers
280
Using LTE
– TDD Signal Analyzer
Procedure
1.
Press the
LIMIT
hot key.
2.
Press the
RF Test Limits
soft key.
3.
Select the test item(s) and set the limit(s) depending on your selected measurement mode:
To set the limit for
Select
Set
Channel power
Channel Power
High Limit
Low Limit
Occupied bandwidth
Occupied BW
High Limit
Spectrum emission mask
Spectrum Emission Mask
(On/Off only)
ACLR
ACLR
(On/Off only)
Multi-ACLR
Multi-ACLR
(On/Off only)
Spurious emissions
Spurious Emissions
(On/Off only)
4.
Optional
. You can enable alarm sound that goes off if the measurement falls outside of the limit.
Toggle the
Beep
soft key between
On
and
Off
to enable or disable the beep sound.
5.
Optional
. Go to
SAVE/LOAD
> Save
, and then select
Limit
to save the limit settings.
See
“Using save” on page 31 for more information.
Conducting power vs. time measurements
The Power vs. Time measurement measures the modulation envelope in the time domain, showing the
signal rise and fall shape of LTE signal and the power of each time slot in an LTE signal.
Power vs. time (slot)
The Power vs. Time (Slot) in LTE - TDD Signal Analyzer measures the modulation envelope in the
time domain, showing the signal rise and fall shapes of LTE signal.
NOTE
In this measurement, desirable level of the input power is lower than -10 dBm. If the input
power to be measured is -10 dBm or higher, it is highly recommended that you use an
external attenuator.
Setting measure setup
After configuring test parameters as described in the
“Configuring test parameters” on page 265, you
can continue your measurement. The measurement settings can be saved and recalled as a file.
Procedure
1.
Press the
MEASURE SETUP
hot key.
2.
To set the nominal channel bandwidth to be measured, complete the following steps:
a.
Press the
Bandwidth
soft key.
b.
Select the bandwidth option:
1.4 MHz
,
3 MHz
,
5 MHz
,
10 MHz
,
15 MHz
, or
20 MHz
.
3.
To select the slot number to be measured, complete the following steps:
Содержание CellAdvisor JD780A Series
Страница 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Страница 91: ...JD780A Series Analyzers Using Spectrum Analyzer 91 Figure 23 PIM detection for multiple carriers...
Страница 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Страница 525: ......
Страница 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...