Using Interference Analyzer
–
To set the limit as the alarm reference, complete the following steps:
a.
Toggle the
Alarm Reference
soft key and select
Line
.
The Limit Line menu becomes activated to be set.
b.
Press the
Limit Line
soft key to define a threshold for the limit line.
c.
Enter a value by using the numeric keys. You can also use the rotary knob.
d.
Press the
dBm
soft key.
6.
Optional
. To adjust the volume for alarm sound, press the
Volume
soft key and then turn the
rotary knob clockwise or counter clockwise.
7.
Optional
. You can save logging data automatically so that you can load the measurement data
and replay in the Spectrum Replayer mode. See
“Using save” on page 31 for more information.
AM/FM audio demodulation
You can use the AM/FM Audio Demodulation to identify if the interfering signal is n
th
harmonics of
AM/FM audio. The instrument sounds demodulated signals that you can hear through the speaker or
headphones.
Procedure
1.
Configure test parameters as described in the
“Configuring test parameters” on page 95.
2.
Press the
MEASURE SETUP
hot key.
3.
Toggle the
Audio Demod
soft key between
On
and
Off
to turn the AM/FM audio demodulation
on and off.
4.
Press the
Demod at
soft key, and then select the marker number for the frequency to be
demodulated.
NOTE
You must set the marker(s) first by using the
MARKER
hard key as the AM/FM
demodulator uses the marker position to demodulate. See
“Using marker” on page 54 for
more information.
5.
Press the
Demod Mode
soft key, and then select the demodulation mode:
AM
,
FM
and
CW
.
6.
Press the
Dwell Time
soft key, and then turn the rotary knob to set the demodulation interval
between 1 and 20.
7.
Press the
Volume
soft key, and then turn the rotary knob to set the speaker volume.
8.
Toggle the
Auto Gain
soft key between
On
and
Off
to enable or disable the automatic gain
feature to adjust the interval gain of the demodulator.
9.
Optional
. Go to
SAVE/LOAD
> Save
, and then select the save option from the choices available
for your measurement mode. See
“Using save” on page 31 for more information.
Interference ID
The Interference ID automatically classifies interfering signals over a designated spectrum and
displays the list of possible signal types corresponding to the selected signal.
Procedure
1.
Configure test parameters as described in the
“Configuring test parameters” on page 95.
2.
Press the
MEASURE SETUP
hot key.
Содержание CellAdvisor JD780A Series
Страница 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Страница 91: ...JD780A Series Analyzers Using Spectrum Analyzer 91 Figure 23 PIM detection for multiple carriers...
Страница 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Страница 525: ......
Страница 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...