Chapter 8
FiberComplete Modules
Activating the function
User Manual
770000102/08
107
Activating the function
1
Press the
H
OME
button.
2
Select the icon FCOMP
.
The icon turns yellow and Fiber Complete is displayed
.
Establishing References
To get a meaningful measurement, the two leads or patch cords used for
the measurement have to be referenced. The references are valid for all
fibers that will be tested during the day with the same patch cords. If, at
anytime, the patchcords are disconnected from the test instruments and/
or have been contaminated by dirt or dust, the patchcords have to be re-
inspected and referencing stage has to be redone.
Fig. 67
Acceptance Test
Pass
Next
Fiber
Measure ORL A & ORL B, and store/display on
both units
Measure Loss A->B, B->A, and Avg and store/display on
both units
Press
S
TART
/S
TOP
button to launch the measurements
Setup Units A & B (Measurement and File parameters)
Reference Source and power meter on Units A & B
Compare results with defined thresholds
Fail
Fault finder
MTS2100 4100 Module US rev08.book Page 107 Jeudi, 1. août 2013 1:43 13