Installation, unpacking

 

4.

Check instrument for any damage caused by shipping. 
If any damage is established, contact the carrier and 
distributor immediately. 
 

Check instrument and any other parts against order. 
 

Installation instructions 

4.1.

Disconnect mains before working on electrical 

part of equipment. 
Keep sufficient clearance around the unit for 

maintenance and the ventilation of the sample 
gas conditioner. 
The equipment has to be connected and 

grounded according to the local rules and regula-
tions. 
The device is designed for use in indoor area. Pay 

attention to adequate ventilation especially when 
installed in closed housings, f.i. analysis cabinets. 
If adequate ventilation, for technical or structural 
reasons is not possible, it is recommended to use 
a forced air cooling, fan or air conditioner, which 
must have no impact on the unit´s ventilation. 

 

Mounting

 

4.2.

Observe mounting position according specification. 

1.

 

Protect unit from excessive exposure to sunlight or 

2.

extreme sources of heat and against rain and dirt. 
Bring sample gas cooler to final mounting position. 

3.

Connect inlet and outlet of the condensate pump 

4.

and check for leaks. 

Connect condensate outlet to condensate collection 

5.

system and check for leaks.

 

Consumables

12.90392

Hose set condensate pump (5 pcs.)

12.90396

Hose set dosing pump (5 pcs.)

K1233007H

Hose set condensate pump Acidflex for 
aggressive condensate (4 pcs.)

K1233009A

Pulley holder condensate pump

K1233011

Tubing cover

Spare parts

K1233014

Pump housing with latch

Accessories

32.90510

JKS-10 condensate collecting vessel, 10 l, 
with floating switch for level monitoring

 

Installation, Sichtkontrolle 

4.

Nach dem Auspacken ist das Gerät auf allfällige Trans-
portschäden zu untersuchen. Wurde ein Schaden fest-
gestellt, sind unverzüglich die verantwortliche Spedition 
und der Händler zu benachrichtigen. 

Es ist zu überprüfen, ob die Gerätelieferung Ihrer Bestel-
lung entspricht. 

Installationsvorschriften 

4.1.

Bei Arbeiten am elektrischen Teil des Gerätes ist 

es vom Netz zu trennen. 
Es ist ausreichender Freiraum um das Gerät, für 

die Wartung, sowie für die Be- und Entlüftung der 
Messgasaufbereitung sicherzustellen. 
Das Gerät muss entsprechend den örtlich gelten-

den Vorschriften angeschlossen und geerdet wer-
den. 
Das Gerät ist für den Einsatz im Innenraumbe-

reich ausgelegt. Dabei ist auf ausreichende Be-
lüftung  insbesondere bei der Montage in 
geschlossenen Gehäusen, wie z.B. Analysen-
schränken zu achten. Ist aus technischen oder 
baulichen Gründen keine ausreichende Belüftung 
möglich wird der Einsatz einer Zwangsbelüftung, 
Ventilator oder Klimagerät empfohlen, was jedoch 
die Gerätebelüftung nicht beeinflussen darf. 

Montage 

4.2.

Einbaulage gemäß der technischen Spezifikation 

1.

beachten. 

Gerät vor direkter Sonnen- oder Wärmestrahlung, 

2.

Regen und starker Verschmutzung schützen. 

Kondensatpumpe in endgültige Montageposition 

3.

bringen. 

Ein- und Ausgangsverbindungen an der Kondensat-

4.

pumpe herstellen und auf Dichtheit überprüfen. 

Kondensatausgang mit Kondensatsammelsystem 

5.

verbinden und auf Dichtheit überprüfen. 

Verschleißteile

12.90392

Schlauchset Kondensatpumpe (5 Stk.)

12.90396

Schlauchset Dosierpumpe (5 Stk.)

K1233007H

Schlauchset Kondensatpumpe Acidflex für 
aggressive Kondensate (4 Stk.)

K1233009A

Rollenträger Kondensatpumpe

K1233011

Laufband

Ersatzteile

K1233014

Pumpengehäuse mit Drehriegel

Zubehör

32.90510

JKS-10 Kondensat-Sammelbehälter, 10 l, mit 
Schwimmerschalter für Füllstandsüberwa-
chung

Manual JSR-25

BA_DE_JSR-25_v1.2                                              

 ––––––––––––––– 

[ 5 / 14 ]

 –––––––––––––––––

Содержание JSR-25

Страница 1: ...JCT ANALYSENTECHNIK GMBH BEDIENUNGSANLEITUNG OPERATING MANUAL 01 20 JSR 25...

Страница 2: ...d either express or implied including warranties or conditions of merchantability or fitness for a particular purpose Subject to technical modifications without notice Manual JSR 25 BA_DE_JSR 25_v1 2...

Страница 3: ...langer Lagerung unter ung nstigen Verh lt nissen nach schweren Transportbeanspruchungen VORSICHT Vor Entfernen der Gasanschl sse oder von Bauteilen der Kondensatableitung ist die Messgaszufuhr abzu s...

Страница 4: ...5 230 VAC 50 60 Hz 1 U min 12ml h 32 00550 JSR 25 Dosierpumpe ohne Geh use 115 230 VAC 50 60 Hz 1 U min 12ml h BA_DE_JSR 25_v1 2 4 14 Modell JSR 25 JSR 25 im Geh use Beschreibung Kontinuierlich arbeit...

Страница 5: ...olle 4 Nach dem Auspacken ist das Ger t auf allf llige Trans portsch den zu untersuchen Wurde ein Schaden fest gestellt sind unverz glich die verantwortliche Spedition und der H ndler zu benachrichtig...

Страница 6: ...4 3 ACHTUNG Dieses Ger t wird mit Netzspannung betrieben Beim Betrieb stehen zwangsl ufig bestimmte Teile unter ge f hrlicher Spannung Durch Abnahme des Geh uses werden spannungsf h rende Teile zug n...

Страница 7: ...ce Manual JSR 25 Inbetriebnahme 5 VORSICHT Vor dem Einschalten ist sicherzustellen dass die am Ger t eingestellte Betriebsspannung und die Netzspan nung bereinstimmen 1 Dichtheitspr fung der Verbindun...

Страница 8: ...un denen Gefahren vertraut ist darf Reperatur und Ein stellarbeiten am offenen und angeschlossenen Ger t vornehmen Kondensatpumpe 6 1 Pumpenschlauch Rollentr ger und Laufband der Kon densatpumpe sind...

Страница 9: ...sate pump tubes from fittings Schritt 3 Step 3 Verschlussriegel im Uhrzeigersinn ff nen und Laufband mit Schlauchein satz entfernen Remove tubing cover including the pump tube by rotation of lock clip...

Страница 10: ...s in the rail until they snap in Schritt 6 Step 6 Verschlussriegel gegen den Uhrzei gersinn drehen bis dieser einrastet Twist back the lock clip counter clock wise until both ends snap in properly Sch...

Страница 11: ...ger eingelaufen Rollentr ger tauschen Malfunction Cause remedy Wet sample gas Malfunction of condensate pump Replace tube set call JCT service Inefficient condensate removal Replace tube set Pay atte...

Страница 12: ...ons in mm Manual JSR 25 Abmessungen 9 JSR 25 ohne Geh use JSR 25 mit Geh use Alle Abma e in mm BA_DE_JSR 25_v1 2 12 14 B 8 86 0 8 68 0 98 98 8 Analysentechnik GmbH A 2700 Wiener Neustadt phone 43 2622...

Страница 13: ...Manual JSR 25 BA_DE_JSR 25_v1 2 13 14...

Страница 14: ...der sterr Sparkassen AG BIC GIBAATWWXXX IBAN AT25 2011 1829 4301 5600 UNICREDIT BANK AUSTRIA AG BIC BKAUATWW IBAN AT89 1100 0039 9372 2200 UID Nr VAT No ATU20559900 Kreisgericht Wr Neustadt FN 119682...

Отзывы: